Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man
297,963 chars · English
The Project Gutenberg eBook of Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man
This eBook is for the use of anyone anywhere in the United States and
most other parts of the world at no cost and with almost no restrictions
whatsoever. You may copy it, give it away or re-use it under the terms
of the Project Gutenberg License included with this eBook or online
at www.gutenberg.org. If you are not located in the United States,
you will have to check the laws of the country where you are located
before using this eBook.
Title: Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man
Author: Francis Gano Benedict
Thorne M. Carpenter
Release date: February 28, 2009 [eBook #28216]
Most recently updated: January 4, 2021
Language: English
Other information and formats: www.gutenberg.org/ebooks/28216
Credits: Produced by Bryan Ness, Josephine Paolucci and the Online
Distributed Proofreading Team at https://www.pgdp.net. (This
book was produced from scanned images of public domain
material from the Google Print project.)
*** START OF THE PROJECT GUTENBERG EBOOK RESPIRATION CALORIMETERS FOR STUDYING THE RESPIRATORY EXCHANGE AND ENERGY TRANSFORMATIONS OF MAN ***
Produced by Bryan Ness, Josephine Paolucci and the Online
Distributed Proofreading Team at https://www.pgdp.net. (This
book was produced from scanned images of public domain
material from the Google Print project.)
Respiration Calorimeters for Studying the Respiratory Exchange and
Energy Transformations of Man
BY
FRANCIS G. BENEDICT and THORNE M. CARPENTER
[Illustration]
WASHINGTON, D. C.
PUBLISHED BY THE CARNEGIE INSTITUTION OF WASHINGTON
1910
CARNEGIE INSTITUTION OF WASHINGTON
PUBLICATION NO. 123
The Lord Baltimore Press
BALTIMORE, MD., U. S. A.
PREFACE.
The immediate development and construction of suitable apparatus for
studying the complicated processes of metabolism in man was obviously
the first task in equipping the Nutrition Laboratory. As several series
of experiments have already been made with these respiration
calorimeters, it is deemed advisable to publish the description of the
apparatus as used at present. New features in the apparatus are,
however, frequently introduced as opportunity to increase accuracy or
facilitate manipulation is noted.
We wish here to express our sense of obligation to the following
associates: Mr. W. E. Collins, mechanician of the Nutrition Laboratory,
constructed the structural steel framework and contributed many
mechanical features to the apparatus as a whole; Mr. J. A. Riche,
formerly associated with the researches in nutrition in the chemical
laboratory of Wesleyan University, added his previous experience in
constructing and installing the more delicate of the heating and cooling
devices. Others who have aided in the painstaking construction, testing,
and experimenting with the apparatus are Messrs. W. H. Leslie, L. E.
Emmes, F. L. Dorn, C. F. Clark, F. A. Renshaw, H. A. Stevens, Jr., Miss
H. Sherman, and Miss A. Johnson.
The numerous drawings were made by Mr. E. H. Metcalf, of our staff.
BOSTON, MASSACHUSETTS,
_August 10, 1909._
CONTENTS.
PAGE
Introduction 1
Calorimeter laboratory 3
General plan of calorimeter laboratory 3
Heating and ventilating 7
The calorimeter 10
Fundamental principles of the apparatus 10
The calorimeter chamber 11
General construction 14
Prevention of radiation 17
The thermo-electric elements 19
Interior of the calorimeter 20
Heat-absorbing circuit 22
Thermometers 26
Mercurial thermometers 26
Electric-resistance thermometers 28
Air-thermometers 28
Wall thermometers 29
Electrical rectal thermometer 29
Electric-resistance thermometers for the water-current 29
Observer's table 31
Connections to thermal-junction systems 33
Rheostat for heating 34
Wheatstone bridges 34
Galvanometer 35
Resistance for heating coils 35
Temperature recorder 36
Fundamental principle of the apparatus 38
The galvanometer 39
The creeper 40
The clock 42
Installation of the apparatus 42
Temperature control of the ingoing air 43
The heat of vaporization of water 44
The bed calorimeter 45
Measurements of body-temperature 48
Control experiments with the calorimeter 50
Determination of the hydrothermal equivalent of the calorimeter 52
General description of the respiration apparatus 54
Testing the chamber for tightness 54
Ventilation of the chamber 54
Openings in the chamber 55
Ventilating air-current 57
Blower 57
Absorbers for water-vapor 58
Potash-lime cans 60
Balance for weighing absorbers 61
Purification of the air-current with sodium bicarbonate 63
Valves 63
Couplings 64
Absorber table 65
Oxygen supply 67
Automatic control of oxygen supply 69
Tension equalizer 71
Barometer 72
Analysis of residual air 73
Gas-meter 75
Calculation of results 76
Analysis of oxygen 76
Advantage of a constant-temperature room and temperature
control 77
Variations in the apparent volume of air 77
Changes in volume due to the absorption of water and
carbon dioxide 78
Respiratory loss 78
Calculation of the volume of air residual in the chamber 79
Residual analyses 80
Calculation from residual analyses 80
Influence of fluctuations in temperature and pressure
on the apparent volume of air in the system 83
Influence of fluctuations in the amounts of carbon
dioxide and water-vapor upon residual oxygen 83
Control of residual analyses 84
Nitrogen admitted with the oxygen 84
Rejection of air 85
Interchange of air in the food aperture 85
Use of the residual blank in the calculations 86
Abbreviated method of computation of oxygen admitted to
the chamber for use during short experiments 88
Criticism of the method of calculating the volume of oxygen 89
Calculation of total output of carbon dioxide and
water-vapor and oxygen absorption 91
Control experiments with burning alcohol 91
Balance for weighing subject 93
Pulse rate and respiration rate 95
Routine of an experiment with man 96
Preparation of subject 96
Sealing in the cover 97
Routine at observer's table 97
Manipulation of the water-meter 98
Absorber table 99
Supplemental apparatus 100
ILLUSTRATIONS.
PAGE
Fig. 1. General plan of respiration calorimeter laboratory 4
2. General view of laboratory taken near main door 4
3. General view of laboratory taken near refrigeration room 4
4. General view of laboratory taken near temperature recorder 4
5. View of laboratory taken from entrance of bed calorimeter 4
6. Plan of heating and ventilating the calorimeter laboratory 6
7. Horizontal cross-section of chair calorimeter 11
8. Vertical cross-section of chair calorimeter 12
9. Vertical cross-section of chair calorimeter from front to back 13
10. Photograph of framework of chair calorimeter 14
11. Photograph of portion of framework and copper shell 14
12. Cross-section in detail of walls of calorimeter 16
13. Detail of drop-sight feed-valve and arrangement of outside
cooling circuit 18
14. Schematic diagram of water-circuit for the heat-absorbers of
the calorimeter 22
15. Detail of air-resistance thermometer 28
16. Details of resistance thermometers for water-circuit 30
17. Diagram of wiring of observer's table 32
18. Diagram of rheostat and resistances in series with it 36
19. Diagram of wiring of differential circuit with shunts used
with resistance thermometers for water-circuit 38
20. Diagram of galvanometer coil, used with recording apparatus
for resistance thermometers in water-circuit 40
21. Diagram of wiring of circuits actuating plunger and creeper 41
22. Diagram of wiring of complete 110-volt circuit 41
23. Temperature recorder 42
24. Detailed wiring diagram showing all parts of the recording
apparatus, together with wiring to thermometers 42
25. Section of calorimeter walls and portion of ventilating
air-circuit 43
26. Cross-section of bed calorimeter 46
27. Diagram of ventilation of the respiration calorimeter 57
28. Cross-section of sulphuric acid absorber 59
29. Balance for weighing absorbers 62
30. Diagram of absorber table 66
31. Diagram of oxygen balance and cylinders 68
32. The oxygen cylinder and connections to tension equalizer 70
RESPIRATION CALORIMETERS FOR STUDYING THE RESPIRATORY EXCHANGE AND
ENERGY TRANSFORMATIONS IN MAN.
INTRODUCTION.
The establishment in Boston of an inquiry into the nutrition of man with
the construction of a special laboratory for that purpose is a direct
outcome of a series of investigations originally undertaken in the
chemical laboratory of Wesleyan University, in Middletown, Connecticut,
by the late Prof. W. O. Atwater. Appreciating the remarkable results of
Pettenkofer and Voit[1] and their associates, as early as 1892 he made
plans for the construction of a respiration apparatus accompanied by
calorimetric features. The apparatus was designed on the general
ventilation plan of the above investigators, but in the first
description of this apparatus[2] it is seen that the method used for the
determination of carbon dioxide and water-vapor was quite other than
that used by Voit. Each succeeding year of active experimenting brought
about new developments until, in 1902, the apparatus was essentially
modified by changing it from the open-circuit type to the closed-circuit
type of Regnault and Reiset. This apparatus, thus modified, has been
completely described in a former publication.[3] The calorimetric
features likewise underwent gradual changes and, as greater accuracy was
desired, it was found impracticable to conduct calorimetric
investigations to the best advantage in the basement of a chemical
laboratory. With four sciences crowded into one building it was
practically impossible to devote more space to these researches.
Furthermore, the investigations had proceeded to such an extent that it
seemed desirable to construct a special laboratory for the purpose of
carrying out the calorimetric and allied investigations on the nutrition
of man.
In designing this laboratory it was planned to overcome the difficulties
experienced in Middletown with regard to control of the room-temperature
and humidity, and furthermore, while the researches had heretofore been
carried on simultaneously with academic duties, it appeared absolutely
necessary to adjust the research so that the uninterrupted time of the
experimenters could be given to work of this kind. Since these
experiments frequently continued from one to ten days, their
satisfactory conduct was not compatible with strenuous academic duties.
As data regarding animal physiology began to be accumulated, it was soon
evident that there were great possibilities in studying abnormal
metabolism, and hence the limited amount of pathological material
available in Middletown necessitated the construction of the laboratory
in some large center.
A very careful consideration was given to possible sites in a number of
cities, with the result that the laboratory was constructed on a plot of
ground in Boston in the vicinity of large hospitals and medical schools.
Advantage was taken, also, of the opportunity to secure connections with
a central power-plant for obtaining heat, light, electricity, and
refrigeration, thus doing away with the necessity for private
installation of boilers and electrical and refrigerating machinery. The
library advantages in a large city were also of importance and within a
few minutes' walk of the present location are found most of the large
libraries of Boston, particularly the medical libraries and the
libraries of the medical schools.
The building, a general description of which appeared in the Year Book
of the Carnegie Institution of Washington for 1908, is of plain brick
construction, trimmed with Bedford limestone. It consists of three
stories and basement and practically all the space can be used for
scientific work. Details of construction may be had by reference to the
original description of the building. It is necessary here only to state
that the special feature of the new building with which this report is
concerned is the calorimeter laboratory, which occupies nearly half of
the first floor on the northern end of the building.
FOOTNOTES:
[1] Pettenkofer and Voit: Ann. der Chem. u. Pharm. (1862-3), Supp. Bd.
2, p. 17.
[2] Atwater, Woods, and Benedict: Report of preliminary investigations
on the metabolism of nitrogen and carbon in the human organism with a
respiration calorimeter of special construction, U. S. Dept. of Agr.,
Office of Experiment Stations Bulletin 44. (1897.)
[3] W. O. Atwater and F. G. Benedict: A respiration calorimeter with
appliances for the direct determination of oxygen. Carnegie Institution
of Washington Publication No. 42. (1905.)
CALORIMETER LABORATORY.
The laboratory room is entered from the main hall by a double door. The
room is 14.2 meters long by 10.1 meters wide, and is lighted on three
sides by 7 windows. Since the room faces the north, the temperature
conditions are much more satisfactory than could be obtained with any
other exposure. In constructing the building the use of columns in this
room was avoided, as they would interfere seriously with the
construction of the calorimeters and accessory apparatus. Pending the
completion of the five calorimeters designed for this room a temporary
wooden floor was laid, thus furnishing the greatest freedom in placing
piping and electric wiring beneath the floor. As fast as the
calorimeters are completed, permanent flooring with suitably covered
trenches for pipes is to be laid. The room is amply lighted during the
day, the windows being very high, with glass transoms above. At night a
large mercury-vapor lamp in the center of the room, supplemented by a
number of well-placed incandescent electric lights, gives ample
illumination.
GENERAL PLAN OF CALORIMETER LABORATORY.
The general plan of the laboratory and the distribution of the
calorimeters and accessory apparatus are shown in fig. 1. The double
doors lead from the main hall into the room. In general, it is planned
to conduct all the chemical and physical observations as near the center
of the laboratory as possible, hence space has been reserved for
apparatus through the center of the room from south to north. The
calorimeters are on either side. In this way there is the greatest
economy of space and the most advantageous arrangement of apparatus.
At present two calorimeters are completed, one under construction, and
two others are planned. The proposed calorimeters are to be placed in
the spaces inclosed by dotted lines. Of the calorimeters that are
completed, the so-called chair calorimeter, which was the first built,
is in the middle of the west side of the room, and immediately to the
north of it is the bed calorimeter, already tested and in actual use. On
the east side of the room it is intended to place large calorimeters,
one for continuous experiments extending over several days and the other
large enough to take in several individuals at once and to have
installed apparatus and working machinery requiring larger space than
that furnished by any of the other calorimeters. Near the chair
calorimeter a special calorimeter with treadmill is shortly to be built.
The heat insulation of the room is shown by the double windows and the
heavy construction of the doors other than the double doors. On entering
the room, the two calorimeters are on the left, and, as arranged at
present, both calorimeters are controlled from the one platform, on
which, is placed the observer's table, with electrical connections and
the Wheatstone bridges for temperature measurements; above and behind
the observer's table are the galvanometer and its hood. At the left of
the observer's platform is a platform scale supporting the water-meter,
with plug valve and handle conveniently placed for emptying the meter.
The absorption system is placed on a special table conveniently situated
with regard to the balance for weighing the absorbers. The large balance
used for weighing the oxygen cylinders is directly across the center
aisle and the analytical balance for weighing the U-tubes for residual
analysis is near by.
[Illustration: FIG. 1.--General plan of respiration calorimeter
laboratory.]
[Illustration: FIG. 2
General view of laboratory room taken near the main door. At the extreme
right is the absorber table, and back of it the bed calorimeter. In the
immediate foreground is shown the balance for weighing absorbers. A
sulphuric acid absorber is suspended on the left hand arm of the
balance. At the left is the observer's table and back of it the chair
calorimeter with a large balance above for weighing subjects. On the
floor, to the left, is the water meter for weighing water used to bring
away heat.]
[Illustration: FIG. 3
General view of laboratory taken near the refrigeration room. The
observer's table is in the immediate foreground with water balance at
the left, and chair calorimeter with balance for weighing man at the
extreme left. At the right of the observer's table is the absorption
system table, and on the wall in the rear the temperature recorder. At
the right is shown the balance for weighing absorbers, and back of that
the case surrounding the balance for weighing oxygen.]
[Illustration: FIG. 4
General view of laboratory taken near the temperature recorder. The bed
calorimeter is at the right, the absorber table in the immediate
foreground, back of it the chair calorimeter and observer's table, and
at the left the balance for weighing absorbers. Near the ceiling are
shown the ducts for the cold air used for temperature control.]
[Illustration: FIG. 5
View of laboratory taken from the entrance of the bed calorimeter, with
balance for weighing oxygen cylinders at the left. The structural steel
skeleton of the calorimeter for long experiments is at the right and
sections of the copper lining are in the rear, resting against the
wall.]
Another view of the laboratory, taken near the door leading to the
refrigeration room, is shown in fig. 3. At the right is seen the balance
used for weighing absorbers, and back of it, imperfectly shown, is the
case surrounding the balance for weighing oxygen cylinders. On the wall,
in the rear, is the recording apparatus for electric resistance
thermometers in the water-circuit, a detail of which is shown in fig.
23. In the foreground in the center is seen the observer's table; at the
right of this is shown the table for the absorption system, and at the
left the chair calorimeter with the balance for weighing subjects above
it. The mercury-vapor light, which is used to illuminate the room, is
immediately above the balance for weighing absorbers.
[Illustration: FIG. 6.--Plan of heating and ventilating calorimeter
laboratory, showing general plan of circulation of the special cooling
system and the position of the thermostats and radiators which they
control. The two small diagrams are cross-sections of brine and heating
coils.]
The bed calorimeter and the absorbing-system table are better shown in
fig. 4, a general view of the laboratory taken near the temperature
recorder. In the immediate foreground is the table for the absorption
system, and back of it are the observer's table and chair calorimeter.
At the right, the bed calorimeter with the front removed and the rubber
hose connections as carried from the absorber table to the bed
calorimeter are shown. At the extreme left is the balance for weighing
the absorbers. Above the chair calorimeter can be seen the balance for
weighing the subject, and at its right the galvanometer suspended from
the ceiling.
The west side of the laboratory at the moment of writing contains the
larger proportion of the apparatus. On the east side there exist only
the balance for weighing oxygen cylinders and an unfinished[4] large
calorimeter, which will be used for experiments of long duration. A view
taken near the front end of the bed calorimeter is shown in fig. 5. At
the right, the structural skeleton of the large calorimeter is clearly
shown. Some of the copper sections to be used in constructing the lining
of the calorimeter can be seen against the wall in the rear.
At the left the balance for weighing the oxygen cylinders is shown with
its counterpoise. A reserve oxygen cylinder is standing immediately in
front of it. A large calorimeter modeled somewhat after the plan of
Sondén and Tigerstedt's apparatus in Stockholm and Helsingfors is
planned to be built immediately back of the balance for weighing oxygen
cylinders.
HEATING AND VENTILATING.
Of special interest in connection with this calorimeter laboratory are
the plans for maintaining constant temperature and humidity (fig. 6).
The room is heated by five steam radiators (each with about 47 square
feet of radiating surface) placed about the outer wall, which are
controlled by two pendant thermostats. A certain amount of indirect
ventilation is provided, as indicated by the arrows on the inner wall.
The room is cooled and the humidity regulated by a system of
refrigeration installed in an adjoining room. This apparatus is of
particular interest and will be described in detail.
In the small room shown at the south side of the laboratory is placed a
powerful electric fan which draws the air from above the floor of the
calorimeter laboratory, draws it over brine coils, and sends it out into
a large duct suspended on the ceiling of the laboratory. This duct has a
number of openings, each of which can be controlled by a valve, and an
unlimited supply of cold air can be directed to any portion of the
calorimeter room at will. To provide for more continuous operation and
for more exact temperature control, a thermostat has been placed in the
duct and is so constructed as to operate some reheater coils beneath the
brine-coils in the refrigerating room. This thermostat is set at 60° F.,
and when the temperature of the air in the duct falls below this point,
the reheater system is automatically opened or closed. The thermostat
can be set at any point desired. Up to the present time it has been
unnecessary to utilize this special appliance, as the control by hand
regulation has been most satisfactory.
Two vertical sections through the refrigerating coils are shown in fig.
6. Section A-B shows the entrance near the floor of the calorimeter
room. The air is drawn down over the coils, passes through the blower,
and is forced back again to the top of the calorimeter room into the
large duct. If outdoor air is desired, a special duct can be connected
with the system so as to furnish outdoor air to the chamber. This has
not as yet been used. Section C-D shows the fan and gives a section
through the reheater. The brine coils, 400 meters long, are in
triplicate. If one set becomes covered with moisture and is somewhat
inefficient, this can be shut off and the other two used. When the
frozen moisture melts and drops off, the single coil can be used again.
It has been found that the system so installed is most readily
controlled.
The degree of refrigeration is varied in two ways: (1) the area of brine
coils can be increased or decreased by using one, two, or all three of
the coils; or (2) the amount of air passing over the cooling pipes may
be varied by changing the speed of the blower. In practice substantially
all of the regulation is effected by varying the position of the
controlling lever on the regulating rheostat. The apparatus functionates
perfectly and the calorimeter room can be held at 20° C. day in and day
out, whether the temperature outdoors is 40° below or 100° above 0° F.
It can be seen, also, that this system provides a very satisfactory
regulation of the humidity, for as the air passes over the brine coils
the moisture is in large part frozen out. As yet, no hygrometric study
has been made of the air conditions over a long period, but the
apparatus is sufficiently efficient to insure thorough electrical
insulation and absence of leakage in the intricate electrical
connections on the calorimeters.
The calorimeters employ the thermo-electric element with its low
potential and a D'Arsonval galvanometer of high sensibility, and in
close proximity it is necessary to use the 110-volt current for heating,
consequently the highest degree of insulation is necessary to prevent
disturbing leakage of current.
The respiration calorimeter laboratory is so large, the number of
assistants in the room at any time is (relatively speaking) so small,
seldom exceeding ten, and the humidity and temperature are so very
thoroughly controlled, that as yet it has been entirely unnecessary to
utilize even the relatively small amount of indirect ventilation
provided in the original plans.
During the greater part of the winter it is necessary to use only one of
the thermostats and the radiators connected with the other can be shut
off, since each radiator can be independently closed by the valves on
the steam supply and return which go through the floor to the basement.
The temperature control of this room is therefore very satisfactory and
economical.
It is not necessary here to go into the advantages of temperature
control of the working rooms during the summer months. Every one seems
to be thoroughly convinced that it is necessary to heat rooms in the
winter, but our experience thus far has shown that it is no less
important to cool the laboratory and control the temperature and
moisture during the summer months, as by this means both the efficiency
and endurance of the assistants, to say nothing of the accuracy of the
scientific measurements, are very greatly increased. Arduous scientific
observations that would be wholly impossible in a room without
temperature control can be carried on in this room during the warmest
weather.
FOOTNOTES:
[4] As this report goes to press, this calorimeter is well on the way to
completion.
THE CALORIMETER.
In describing this apparatus, for the sake of clearness, the
calorimetric features will be considered before the appliances for the
determination of the respiratory products.
FUNDAMENTAL PRINCIPLES OF THE APPARATUS.
The measurements of heat eliminated by man, as made by this apparatus,
are based upon the fact that the subject is inclosed in a heat-proof
chamber through which a current of cold water is constantly passing. The
amount of water, the flow of which, for the sake of accuracy, is kept at
a constant rate, is carefully weighed. The temperatures of the water
entering and leaving the chamber are accurately recorded at frequent
intervals. The walls of the chamber are held adiabatic, thus preventing
a gain or loss of heat by arbitrarily heating or cooling the outer metal
walls, and the withdrawal of heat by the water-current is so controlled,
by varying the temperature of the ingoing water, that the heat brought
away from the calorimeter is exactly equal in amount to the heat
eliminated by radiation and conduction by the subject, thus maintaining
a constant temperature inside of the chamber. The latent heat of the
water vaporized is determined by measuring directly the water vapor in
the ventilating air-current.
In the construction of the new calorimeters a further and fundamental
change in construction has been made in that all the thermal junctions,
heating wires, and cooling pipes have been attached directly to the zinc
wall of the calorimeter, leaving the outer insulating panels free from
incumbrances, so that they can be removed readily and practically all
parts inspected whenever desired without necessitating complete
dismantling of the apparatus. This arrangement is possible except in
those instances where connections pass clear through from the interior
of the chamber to the outside, namely, the food-aperture, air-pipes,
water-pipes, electrical connections, and tubes for connections with
pneumograph and stethoscope; but the apparatus is so arranged as to have
all of these openings in one part of the calorimeter. It is possible,
therefore, to remove all of the outer sections of the calorimeter with
the exception of panels on the east side.
This fundamental change in construction has proven highly advantageous.
It does away with the necessity of rolling the calorimeter out of its
protecting insulating house and minimizes the delay and expense
incidental to repairs or modifications. As the calorimeter is now
constructed, it is possible to get at all parts of it from the outside,
with the exception of one small fixed panel through which the above
connections are passed. This panel, however, is made as narrow as
possible, so that practically all changes can be made by taking out the
adjacent panels.
THE CALORIMETER CHAMBER.
[Illustration: FIG. 7.--Horizontal cross-section of chair calorimeter,
showing cross-section of copper wall at A, zinc wall at B, hair-felt at
E, and asbestos outer wall at F; also cross-section of all upright
channels in the steel construction. At the right is the location of the
ingoing and outgoing water and the thermometers. At C is shown the food
aperture, and D is a gasket separating the two parts. The ingoing and
outcoming air-pipes are shown at the right inside the copper wall. The
telephone is shown at the left, and in the center of the drawing is the
chair with its foot-rest, G. In dotted line is shown the opening where
the man enters.]
[Illustration: FIG. 8.--Vertical cross-section of chair calorimeter,
showing part of rear of calorimeter and structural-steel frame. N,
cross-section of bottom horizontal channel supporting asbestos floor J;
H, H, upright channels (at the right is a side upright channel and to
the left of this is an upright rear channel); M horizontal 8-inch
channel supporting calorimeter; Zn, zinc wall; Cu, copper wall; J,
insulating asbestos.]
The respiration chamber used in Middletown, Connecticut, was designed to
permit of the greatest latitude in the nature of the experiments to be
made with it. As a result, it was found at the end of a number of years
of experimenting that this particular size of chamber was somewhat too
small for the most satisfactory experiments during muscular work and, on
the other hand, somewhat too large for the best results during so-called
rest experiments. In the earlier experiments, where no attempt was made
to determine the consumption of oxygen, these disadvantages were not so
apparent, as carbon dioxide could be determined with very great
accuracy; but with the attempts to measure the oxygen it was found that
the large volume of residual air inside the chamber, amounting to some
4,500 liters, made possible very considerable errors in this
determination, for, obviously, the subject could draw upon the oxygen
residual in the air of the chamber, nearly 1,000 liters, as well as upon
the oxygen furnished from outside sources. The result was that a very
careful analysis of the residual air must be made frequently in order to
insure that the increase or decrease in the amount of oxygen residual in
the air of the chamber was known accurately at the end of each period.
Analysis of this large volume of air could be made with considerable
accuracy, but in order to calculate the exact total of oxygen residual
in the air it was necessary to know the total volume of air inside the
chamber under standard conditions. This necessitated, therefore, a
careful measurement of temperature and pressure, and while the
barometric pressure could be measured with a high degree of accuracy,
it was found to be very difficult to determine exactly the average
temperature of so large a mass of air. The difficulties attending this
measurement and experiments upon this point are discussed in detail
elsewhere.[5] Consequently, as a result of this experience, in planning
the calorimeters for the Nutrition Laboratory it was decided to design
them for special types of experiments. The first calorimeter to be
constructed was one which would have general use in experiments during
rest and, indeed, during experiments with the subject sitting quietly in
the chair.
[Illustration: FIG. 9.--Vertical cross-section of chair calorimeter from
front to back, showing structural steel supporting the calorimeter and
the large balance above for weighing the subject inside the calorimeter.
The chair, method of suspension, and apparatus for raising and lowering
are shown. Part of the heat-absorbers is shown, and their general
direction. The ingoing and outgoing air-pipes and direction of
ventilation are also indicated. The positions of the food-aperture and
wire mat and asbestos support are seen. Surrounding the calorimeter are
the asbestos outside and hair-felt lining.]
It may well be asked why the first calorimeter was not constructed of
such a type as to permit the subject assuming a position on a couch or
sofa, such as is used by Zuntz and his collaborators in their research
on the respiratory exchange, or the position of complete muscular rest
introduced by Johansson and his associates. While the body positions
maintained by Zuntz and Johansson may be the best positions for
experiments of short duration, it was found, as a result of a large
number of experiments, that subjects could be more comfortable and quiet
for periods of from 6 to 8 hours by sitting, somewhat inclined, in a
comfortable arm-chair, provided with a foot-rest. With this in mind the
first calorimeter was constructed so as to hold an arm-chair with a
foot-rest so adjusted that the air-space between the body of the subject
and the walls of the chamber could be cut down to the minimum and thus
increase the accuracy of the determination of oxygen. That the volume
has been very materially reduced may be seen from the fact that the
total volume of the first calorimeter to be described is less than 1,400
liters, or about one-third that of the Middletown apparatus.
GENERAL CONSTRUCTION.
A horizontal cross-section of the apparatus is shown in fig. 7, and a
vertical cross-section facing the front is given in fig. 8. Other
details of structural steel are seen in fig. 9.
In constructing the new chambers, the earlier wood construction, with
its tendency to warp and its general non-rigidity, was avoided by the
use of structural steel, and hence in this calorimeter no use whatever
is made of wood other than the wood of the chair.
To avoid temperature fluctuations due to possible local stratification
of the air in the laboratory, the calorimeter is constructed so as to be
practically suspended in the air, there being a large air-space of some
76 centimeters between the lowest point of the calorimeter and the
floor, and the top of the calorimeter is some 212 centimeters below the
ceiling of the room. Four upright structural-steel channels (4-inch)
were bolted through the floor, so as to secure great rigidity, and were
tied together at the top with structural steel. As a solid base for the
calorimeter chamber two 3-inch channels were placed parallel to each
other 70 centimeters from the floor, joined to these uprights. Upon
these two 3-inch channels the calorimeter proper was constructed. The
steel used for the most part in the skeleton of the apparatus is
standard 2-1/2-inch channel. This steel frame and its support are shown
in fig. 10, before any of the copper lining was put into position. The
main 4-inch channels upon which the calorimeter is supported, the
tie-rods and turn-buckles anchoring the framework to the ceiling, the
I-beam construction at the top upon which is subsequently installed the
large balance for weighing the man, the series of small channels set on
edge upon which the asbestos floor is laid, and the upright row of
channel ribs are all clearly shown.
[Illustration: FIG. 10
Photograph of framework of chair calorimeter. In the photograph are
shown four upright channels and the channels at the top for supporting
the calorimeter. The smaller upright 2-1/2 inch channels and angles are
shown inside of this frame. In the lower part of the figure is seen the
asbestos board for the bottom of the calorimeter and underneath this a
sheet of zinc.]
[Illustration: FIG. 11
Photograph of portion of framework and copper shell. The finished copper
shell is seen in position with some of the thermal junction thimbles
soldered into it. A portion of the food aperture and the four brass
ferrules for conducting the water pipes and air pipes are shown. A
section of the zinc outside is shown in the lower part of the figure.]
A photograph taken subsequently, showing the inner copper lining in
position, is given in fig. 11.
The floor of the chamber is supported by 7 pieces of 2-1/2-inch channel
(N, N, N, fig. 8), laid on top and bolted to the two 3-inch channels (M,
fig. 8). On top of these is placed a sheet of so-called asbestos lumber
(J', fig. 8) 9.5 millimeters thick, cut to fit exactly the bottom of the
chamber. Upright 2-1/2-inch channels (H, fig. 8) are bolted to the two
outside channels on the bottom and to the ends of three of the long
channels between in such a manner as to form the skeleton of the walls.
The upper ends of these channels are fastened together by pieces of
piping (P, P, P, fig. 8) with lock-nuts on either side, thus holding the
whole framework in position.
The I-beams and channels used to tie the four upright channels at the
top form a substantial platform upon which is mounted a large balance
(fig. 9). This platform is anchored to the ceiling at four points by tie
rods and turn-buckles, shown in fig. 4. The whole apparatus, therefore,
is extremely rigid and the balance swings freely.
The top of the chamber is somewhat restricted near the edges (fig. 8)
and two lengths of 2-1/2-inch channel support the sides of the opening
through which the subject enters at the top (fig. 7).
Both the front and back lower channels upon which the bottom rests are
extended so as to provide for supports for the outer walls of asbestos
wood, which serve to insulate the calorimeter. Between the channels
beneath the calorimeter floor and the 3-inch channels is placed a sheet
of zinc which forms the outer bottom metallic wall of the chamber.
In order to prevent conduction of heat through the structural steel all
contact between the inner copper wall and the steel is avoided by having
strips of asbestos lumber placed between the steel and copper. These are
shown as J in fig. 8 and fig. 12. A sheet of asbestos lumber beneath the
copper bottom likewise serves this purpose and also serves to give a
solid foundation for the floor. The supporting channels are placed near
enough together to reinforce fully the sheet of asbestos lumber and
enable it to support solidly the weight of the man. The extra strain on
the floor due to tilting back a chair and thus throwing all the weight
on two points was taken into consideration in planning the asbestos and
the reinforcement by the steel channels. The whole forms a very
satisfactory flooring.
_Wall construction and insulation._--The inner wall of the chamber
consists of copper, preferably tinned on both sides, thus aiding in
soldering, and the tinned inner surface makes the chamber somewhat
lighter. Extra large sheets are obtained from the mill, thus reducing to
a minimum the number of seams for soldering, and seams are made tight
only with difficulty. The copper is of standard gage, the so-called
14-ounce copper, weighing 1.1 pounds per square foot or 5.5 kilograms
per square meter. It has a thickness of 0.5 millimeter. The whole
interior of the skeleton frame of the structural steel is lined with
these sheets; fig. 11 shows the copper shell in position.
For the outer metallic wall, zinc, as the less expensive metal, is used.
One sheet of this material perforated with holes for the attachment of
bolts and other appliances is shown in position on the outside of the
wall in fig. 11. The sheet zinc of the floor is obviously put in
position before the channels upon which it rests are laid. The zinc is
obtained in standard size, and is the so-called 9-ounce zinc, or 0.7
pound to the square foot, or 3.5 kilograms to the square meter. The
sheet has a thickness of 0.5 millimeter.
[Illustration: FIG. 12.--Cross-section in detail of walls of
calorimeter, showing zinc and copper walls and asbestos outside (A);
hair-felt lining (B); cross-section of channel iron (H); brass washer
soldered to copper (K); asbestos insulation between channel iron and
copper (J); bolt holding the whole together (I); heating wire (W) and
insulator holding it (F) shown in air-space between zinc and hair-felt;
section of one of the cooling pipes (C) and its brass support (G);
threaded rod (E) fastened into H at one end and passing through asbestos
wall with a nut on the outside; and iron pipe (D) used as spacer between
asbestos and zinc.]
In the cross-section, fig. 7, A represents the copper wall and B the
zinc wall. Surrounding this zinc wall and providing air insulation is a
series of panels constructed of asbestos lumber, very fire-resisting,
rigid, and light. The asbestos lumber used for these outer panels is 6.4
millimeters (0.25 inch) thick. To further aid in heat insulation we have
glued to the inner face of the different panels a patented material
composed of two layers of sheathing-paper inclosing a half-inch of
hair-felt. This material is commonly used in the construction of
refrigerators. This is shown as E in fig. 7, while the outer asbestos
panels are shown as F.
A detail of the construction of the walls, showing in addition the
heating and cooling devices, is given in fig. 12, in which the copper is
shown held firmly to the upright channel H by means of the bolt I,
screwing into a brass or copper disk K soldered to the copper wall. The
bolt I serves the purpose of holding the copper to the upright channel
and likewise by means of a washer under the head of the screw holds the
zinc to the channel. In order to hold the asbestos-lumber panel A with
the hair-felt lining B a threaded rod E is screwed into a tapped hole in
the outer part of the upright channel H. A small piece of brass or iron
tubing, cut to the proper length, is slipped over this rod and the
asbestos lumber held in position by a hexagonal nut with washer on the
threaded rod E. In this manner great rigidity of construction is
secured, and we have two air-spaces corresponding to the dead air-spaces
indicated in fig. 7, the first between the copper and zinc and the
second between the zinc and hair-felt.
PREVENTION OF RADIATION.
As can be seen from these drawings the whole construction of the
apparatus is more or less of the refrigerator type, _i. e._, there is
little opportunity for radiation or conduction of heat. Such a
construction could be multiplied a number of times, giving a greater
number of insulating walls, and perhaps reducing radiation to the
minimum, but for extreme accuracy in calorimetric investigations it is
necessary to insure the absence of radiation, and hence we have retained
the ingenious device of Rosa, by which an attempt is made arbitrarily to
alter the temperature of the zinc wall so that it always follows any
fluctuations in the temperature of the copper wall. To this end it is
necessary to know _first_ that there is a temperature difference between
zinc and copper and, _second_, to have some method for controlling the
temperature of the zinc. Leaving for a moment the question of measuring
the temperature differences between zinc and copper, we can consider
here the methods for controlling the temperature of the zinc wall.
If it is found necessary to warm the zinc wall, a current of electricity
is passed through the resistance wire W, fig. 12. This wire is
maintained approximately in the middle of the air-space between the zinc
wall and hair-felt by winding it around an ordinary porcelain insulator
F, held in position by a threaded rod screwed into a brass disk soldered
to the zinc wall. A nut on the end of the threaded rod holds the
insulator in position. Much difficulty was had in securing a resistance
wire that would at the same time furnish reasonably high resistance and
would not crystallize or become brittle and would not rust. At present
the best results have been obtained by using enameled manganin wire. The
wire used is No. 28 American wire-gage and has resistance of
approximately 1.54 ohms per foot. The total amount of wire used in any
one circuit is equal to a resistance of approximately 92 ohms. This
method of warming the air-space leaves very little to be desired. It can
be instantaneously applied and can be regulated with the greatest ease
and with the greatest degree of refinement.
If, on the other hand, it becomes necessary to cool the air-space next
to the zinc and in turn cool the zinc, we must resort to the use of cold
water, which is allowed to flow through the pipe C suspended in the
air-space between the zinc and hair-felt at approximately the same
distance as is the heating wire. The support of these pipes is
accomplished by placing them in brass hangers G, soldered to the zinc
and provided with an opening in which the pipe rests.
In the early experimenting, it was found impracticable to use piping of
very small size, as otherwise stoppage as a result of sediment could
easily occur. The pipe found best adapted to the purpose was the
so-called standard one-eighth inch brass pipe with an actual internal
diameter of 7 millimeters. The opening of a valve allowed cold water to
flow through this pipe and the considerable mass of water passing
through produced a very noticeable cooling effect. In the attempt to
minimize the cooling effect of the mass of water remaining in the pipe,
provision was made to allow water to drain out of this pipe a few
moments after the valve was closed by a system of check-valves. In
building the new apparatus, use was made of the compressed-air service
in the laboratory to remove the large mass of cold water in the pipe. As
soon as the water-valve was closed and the air-cock opened, the
compressed air blew all of the water out of the tube.
[Illustration: FIG. 13.--Detail of drop-eight feed-valve and arrangement
of outside cooling circuit. The water enters at A, and the flow is
regulated by the needle-valve at left-hand side. Rate of flow can be
seen at end of exit tube just above the union. The water flows out at C
and compressed air is admitted at B, regulated by the pet-cock.]
The best results have been obtained, however, with an entirely new
principle, namely, a few drops of water are continually allowed to pass
into the pipe, together with a steady stream of compressed air. This
cold water is forcibly blown through the pipe, thus cooling to an amount
regulated by the amount of water admitted. Furthermore, the relatively
dry air evaporates some of the water, thereby producing a somewhat
greater cooling effect. By adjusting the flow of water through the pipe
a continuous cooling effect of mild degree may be obtained. While
formerly the air in the space next the zinc wall was either cooled or
heated alternately by opening the water-valve or by passing a current
through the heating coil, at present it is found much more advantageous
to allow a slow flow of air and water through the pipes continuously,
thus having the air-space normally somewhat cooler than is desired. The
effect of this cooling, therefore, is then counterbalanced by passing an
electric current of varying strength through the heating wire. By this
manipulation it is unnecessary that the observer manipulate more than
one instrument, namely, the rheostat, while formerly he had to
manipulate valves, compressed-air cocks, and rheostat. The arrangement
for providing for the amount of compressed air and water is shown in
fig. 13, in which it is seen that a small drop-sight feed-water valve is
attached to the pipe C leading into the dead air-space surrounding the
calorimeter chamber. Compressed air enters at B and the amount entering
can be regulated by the pet-cock. The amount of water admitted is
readily observed by the sight feed-valve. When once adjusted this form
of apparatus produces a relatively constant cooling effect and
facilitates greatly the manipulation of the calorimetric apparatus as a
whole.
THE THERMO-ELECTRIC ELEMENTS.
In order to detect differences in temperature between the copper and
zinc walls, some system for measuring temperature differences between
these walls is essential. For this purpose we have found nothing that is
as practical as the system of iron-German-silver thermo-electric
elements originally introduced in this type of calorimeter by E. B.
Rosa, of the National Bureau of Standards, formerly professor of physics
at Wesleyan University. In these calorimeters the same principle,
therefore, has been applied, and it is necessary here only to give the
details of such changes in the construction of the elements, their
mounting, and their insulation as have been made as a result of
experience in constructing these calorimeters. An element consisting of
four pairs of junctions is shown in place as T-J in fig. 25.
One ever-present difficulty with the older form of element was the
tendency for the German-silver wires to slip out of the slots in which
they had been vigorously crowded in the hard maple spool. In thus
slipping out of the slots they came in contact with the metal thimble in
the zinc wall and thus produced a ground. In constructing the new
elements four pairs of iron-German-silver thermal junctions were made on
essentially the same plan as that previously described,[6] the only
modification being made in the spool. While the ends of the junctions
nearest the copper are exposed to the air so as to take up most rapidly
the temperature of the copper, it is somewhat difficult to expose the
ends of the junctions nearest the zinc and at the same time avoid
short-circuiting. The best procedure is to extend the rock maple spool
which passes clear through the ferule in the zinc wall and cut a wide
slot in the spool so as to expose the junctions to the air nearest the
ferule. By so doing the danger to the unprotected ends of the junctions
is much less. The two lead-wires of German silver can be carried through
the end of the spool and thus allow the insulation to be made much more
satisfactorily. In these calorimeters free use of these thermal
junctions has been made. In the chair calorimeter there are on the top
16 elements consisting of four junctions each, on the rear 18, on the
front 8, and on the bottom 13. The distribution of the elements is made
with due reference to the direction in which the heat is most directly
radiated and conducted from the surface of the body.
While the original iron-German-silver junctions have been retained in
two of these calorimeters for the practical reason that a large number
of these elements had been constructed beforehand, we believe it will be
more advantageous to use the copper-constantin couple, which has a
thermo-electric force of 40 microvolts per degree as against the 25 of
the iron-German-silver couple. It is planned to install the
copper-constantin junctions in the calorimeters now building.
INTERIOR OF THE CALORIMETER.
Since the experiments to be made with this chamber will rarely exceed 6
to 8 hours, there is no provision made for installing a cot bed or other
conveniences which would be necessary for experiments of long duration.
Aside from the arm-chair with the foot-rest suspended from the balance,
there is practically no furniture inside of the chamber, and a shelf or
two, usually attached to the chair, to support bottles for urine and
drinking-water bottles, completes the furniture equipment. The
construction of the calorimeter is such as to minimize the volume of air
surrounding the subject and yet secure sufficient freedom of movement to
have him comfortable. A general impression of the arrangement of the
pipes, chair, telephone, etc., inside the chamber can be obtained from
figs. 7 and 9. The heat-absorber system is attached to rings soldered to
the ceiling at different points. The incoming air-pipe is carried to the
top of the central dome, while the air is drawn from the calorimeter at
a point at the lower front near the position of the feet of the subject.
From this point it is carried through a pipe along the floor and up the
rear wall of the calorimeter to the exit.
With the perfect heat insulation obtaining, the heat production of the
man would soon raise the temperature to an uncomfortable degree were
there no provisions for withdrawing it. It is therefore necessary to
cool the chamber and, as has been pointed out, the cooling is
accomplished by passing a current of cold water through a heat-absorbing
apparatus permanently installed in the interior of the chamber. The
heat-absorber consists of a continuous copper pipe of 6 millimeters
internal diameter and 10 millimeters external diameter. Along this pipe
there are soldered a large number of copper disks 5 centimeters in
diameter at a distance of 5 millimeters from each other. This increases
enormously the area for the absorption of heat. In order to allow the
absorber system to be removed, added to, or repaired at any time, it is
necessary to insert couplings at several points. This is usually done at
corners where the attachment of disks is not practicable. The total
length of heat-absorbers is 5.6 meters and a rough calculation shows
that the total area of metal for the absorption of heat is 4.7 square
meters. The total volume of water in the absorbers is 254 cubic
centimeters.
It has been found advantageous to place a simple apparatus to mix the
water in the water-cooling circuit at a point just before the water
leaves the chamber. This water-mixer consists of a 15-centimeter length
of standard 1-inch pipe with a cap at each end. Through each of these
caps there is a piece of one-eighth-inch pipe which extends nearly the
whole length of the mixer. The water thus passing into one end returns
inside the 1-inch pipe and leaves from the other. This simple device
insures a thorough mixing.
The air-pipes are of thin brass, 1-inch internal diameter. One of them
conducts the air from the ingoing air-pipe up into the top of the
central dome or hood immediately above the head of the subject. The air
thus enters the chamber through a pipe running longitudinally along the
top of the dome. On the upper side of this pipe a number of holes have
been drilled so as to have the air-current directed upwards rather than
down against the head of the subject. With this arrangement no
difficulties are experienced with uncomfortable drafts and although the
air enters the chamber through this pipe absolutely dry, there is no
uncomfortable sensation of extreme dryness in the air taken in at the
nostrils, nor is the absorption of water from the skin of the face,
head, or neck great enough to produce an uncomfortable feeling of cold.
The other air-pipe, as suggested, receives the air from the chamber at
the lower front and passes around the rear to the point where the
outside air-pipe leaves the chamber.
The chamber is illuminated by a small glass door in the food aperture.
This is a so-called "port" used on vessels. Sufficient light passes
through this glass to enable the subject to see inside the calorimeter
without difficulty and most of the subjects can read with comfort. If an
electric light is placed outside of the window, the illumination is very
satisfactory and repeated tests have shown that no measurable amount of
heat passes through the window by placing a 32 c. p. electric lamp 0.5
meter from the food aperture outside. More recently we have arranged to
produce directly inside the chamber illumination by means of a small
tungsten electric lamp connected to the storage battery outside of the
chamber. This lamp is provided with a powerful mirror and a glass shade,
so that the light is very bright throughout the chamber and is
satisfactory for reading. It is necessary, however, to make a correction
for the heat developed, amounting usually to not far from 3 calories per
hour.
By means of a hand microphone and receiver, the subject can communicate
with the observers outside at will. A push-button and an electric bell
make it possible for him to call the observers whenever desired.
HEAT-ABSORBING CIRCUIT.
To bring away the heat produced by the subject, it is highly desirable
that a constant flow of water of even temperature be secured. Direct
connection with the city supply is not practicable, owing to the
variations in pressure, and hence in constructing the laboratory
building provision was made to install a large tank on the top floor,
fed with a supply controlled by a ball-and-cock valve. By this
arrangement the level in the tank is maintained constant and the
pressure is therefore regular. As the level of the water in the tank is
approximately 9 meters above the opening in the calorimeter, there is
ample pressure for all purposes.
[Illustration: FIG. 14.--Schematic diagram of water circuit for
heat-absorbers of calorimeter. A, constant-level tank from which water
descends to main pipe supplying heat-absorbers; _a_, valve for
controlling supply from tank A; B, section of piping passing into cold
brine; _b_, valve controlling water direct from large tank A; _c_, valve
controlling amount of water from cooling section B; C, thermometer at
mixer; D, electric heater for ingoing water; E, thermometer for ingoing
water; _d d d_, heat-absorbers inside calorimeter; F, thermometer
indicating temperature of outcoming water; G, can for collecting water
from calorimeter; _f_, valve for emptying G.]
The water descends from this tank in a large 2-inch pipe to the ceiling
of the calorimeter laboratory, where it is subdivided into three 1-inch
pipes, so as to provide for a water-supply for three calorimeters used
simultaneously, if necessary, and eliminate the influence of a variation
in the rate of flow in one calorimeter upon the rate of flow in another.
These pipes are brought down the inner wall of the room adjacent to the
refrigeration room and part of the water circuit is passed through a
brass coil immersed in a cooling-tank in the refrigeration room. By
means of a by-pass, water of any degree of temperature from 2° C. to 20°
C. may be obtained. The water is then conducted through a pipe beneath
the floor to the calorimeter chamber, passed through the absorbers, and
is finally measured in the water-meter.
A diagrammatic sketch showing the course of the water-current is given
(fig. 14), in which A is the tank on the top floor controlled by the
ball cock and valve, and _a_ is the main valve which controls this
supply to the cooler B, and by adjusting the valve _b_ and valve _c_
any desired mixture of water can be obtained. A thermometer C gives a
rough idea of the temperature of the water, so as to aid in securing the
proper mixture. The water then passes under the floor of the calorimeter
laboratory and ascends to the apparatus D, which is used for heating it
to the desired temperature before entering the calorimeter. The
temperature of the water as it enters the calorimeter is measured on an
accurately calibrated thermometer E, and it then passes through the
absorber system _d d d_ and leaves the calorimeter, passing the
thermometer F, upon which the final temperature is read. It then passes
through a pipe and falls into a large can G, placed upon scales. When
this can is filled the water is deflected for a few minutes to another
can and by opening valve _f_ the water is conducted to the drain after
having been weighed.
_Brine-tank._--The cooling system for the water-supply consists of a
tank in which there is immersed an iron coil connected by two valves to
the supply and return of the brine mains from the central power-house.
These valves are situated just ahead of the valves controlling the
cooling device in the refrigeration room and permit the passage of brine
through the coil without filling the large coils for the cooling of the
air in the calorimeter laboratory. As the brine passes through this
coil, which is not shown in the figure, it cools the water in which it
is immersed and the water in turn cools the coil through which the
water-supply to the calorimeter passes. The brass coil only is shown in
the figure. The system is very efficient and we have no difficulty in
cooling the water as low as 2° C. As a matter of fact our chief
difficulty is in regulating the supply of brine so as not to freeze the
water-supply.
_Water-mixer._--If the valve _b_ is opened, water flows through this
short length of pipe much more rapidly than through the long coil, owing
to the greater resistance of the cooling coil. In conducting these
experiments the valve c is opened wide and by varying the amount to
which the valve _b_ is opened, the water is evenly and readily mixed.
The thermometer C is in practice immersed in the water-mixer constructed
somewhat after the principle of the mixer inside the chamber described
on page 21. All the piping, including that under the floor, and the
reheater D, are covered with hair-felt and well insulated.
_Rate-valves._--It has been found extremely difficult to secure any form
of valve which, even with a constant pressure of water, will give a
constant rate of flow. In this type of calorimeter it is highly
desirable that the rate of flow be as nearly constant as possible hour
after hour, as this constant rate of flow aids materially in maintaining
the calorimeter at an even temperature. Obviously, fluctuations in the
rate of flow will produce fluctuations in the temperature of the ingoing
water and in the amount of heat brought away. This disturbs greatly the
temperature equilibrium, which is ordinarily maintained fairly constant.
Just before the water enters the reheater D it is caused to pass through
a rate-valve, which at present consists of an ordinary plug-cock. At
present we are experimenting with other types of valves to secure even
greater constancy, if possible.
_Electric reheater._--In order to control absolutely the temperature of
the water entering at E, it is planned to cool the water leaving the
water-mixer at C somewhat below the desired temperature, so that it is
necessary to reheat it to the desired point. This is done by passing a
current of electricity through a coil inserted in the system at the
point D. This electric reheater consists of a standard "Simplex" coil,
so placed in the copper can that the water has a maximum circulation
about the heater. The whole device is thoroughly insulated with
hair-felt. By connecting the electric reheater with the rheostat on the
observer's table, control of the quantity of electricity passing through
the coil is readily obtained, and hence it is possible to regulate the
temperature of the ingoing water to within a few hundredths of a degree.
The control of the amount of heat brought away from the chamber is made
either by (1) increasing the rate of flow or (2) by varying the
temperature of the ingoing water. Usually only the second method is
necessary. In the older form of apparatus a third method was possible,
namely, by varying the area of the absorbing surface of the cooling
system inside of the chamber. This last method of regulation, which was
used almost exclusively in earlier experiments, called for an elaborate
system of shields which could be raised or lowered at will by the
operator outside, thus involving an opening through the chamber which
was somewhat difficult to make air-tight and also considerably
complicating the mechanism inside the chamber. The more recent method of
control by regulating the temperature of the ingoing water by the
electric reheater has been much refined and has given excellent service.
_Insulation of water-pipes through the wall._--To insulate the
water-pipes as they pass through the metal walls of the calorimeter and
to prevent any cooling effect not measured by the thermometers presented
great difficulties. The device employed in the Middletown chamber was
relatively simple, but very inaccessible and a source of more or less
trouble, namely, a large-sized glass tube embedded in a large round
wooden plug with the annular space between the glass and wood filled
with wax. An attempt was made in the new calorimeters to secure air
insulation by using a large-sized glass tube, some 15 millimeters
internal diameter, and passing it through a large rubber stopper,
fitting into a brass ferule soldered between the zinc and copper walls.
(See N, fig. 25.) So far as insulation was concerned, this arrangement
was very satisfactory, but unfortunately the glass tubes break readily
and difficulty was constantly experienced. An attempt was next made to
substitute hard-rubber tubing for the glass tube, but this did not prove
to be an efficient insulator. More recently we have used with perfect
success a special form of vacuum-jacketed glass tube, which gives the
most satisfactory insulation. However, this system of insulation is
impracticable when electric-resistance thermometers are used for
recording the water-temperature differences and can be used only when
mercurial thermometers exclusively are employed. The electric-resistance
thermometers are constructed in such a way, however, as to make
negligible any inequalities in the passage of heat through the
hard-rubber casing. This will be seen in the discussion of these
thermometers.
_Measuring the water._--As the water leaves the respiration chamber it
passes through a valve which allows it to be deflected either into the
drain during the preliminary period, or into a small can where the
measurements of the rate of flow can readily be made, or into a large
tank (G, fig. 14) where the water is weighed. The measurement of the
water is made by weight rather than by volume, as it has been found that
the weighing may be carried out with great accuracy. The tank, a
galvanized-iron ash-can, is provided with a conical top, through an
opening in which a funnel is placed. The diagram shows the water leaving
the calorimeter and entering the meter through this funnel, but in
practice it is adjusted to enter through an opening on the side of the
meter. After the valve _f_ is tightly closed the empty can is weighed.
When the experiment proper begins the water-current is deflected so as
to run into this can and at the end of an hour the water is deflected
into a small can used for measuring the rate of flow. While it is
running into this can, the large can G is weighed on platform scales to
within 10 grams. After weighing, the water is again deflected into the
large can and that collected in the small measuring can is poured into G
through the funnel. The can holds about 100 liters of water and
consequently from 3 to 8 one-hour periods, depending upon the rate of
flow, can be continued without emptying the meter. When it is desired to
empty the meter at the end of the period, the water is allowed to flow
into the small can, and after weighing G, the valve _f_ is opened. About
4 minutes are required to empty the large can. After this the valve is
again closed, the empty can weighed, and the water in the small
measuring-can poured into the large can G through the funnel. The scales
used are the so-called silk scales and are listed by the manufacturers
to weigh 150 kilograms. This form of scales was formerly used in
weighing the man inside the chamber.[7]
THERMOMETERS.
In connection with the calorimeter and the accessories, mercurial and
electric-resistance thermometers are employed. For measuring the
temperature of the water as it enters and leaves the chamber through
horizontal tubes, mercurial thermometers are used, and these are
supplemented by electric-resistance thermometers which are connected
with a special form of recording instrument for permanently recording
the temperature differences. For the measurement of the temperatures
inside of the calorimeter, two sets of electric-resistance thermometers
are used, one of which is a series of open coils of wire suspended in
the air of the chamber so as to take up quickly the temperature of the
air. The other set consists of resistance coils encased in copper boxes
soldered to the copper wall and are designed to indicate the temperature
of the copper wall rather than that of the air.
MERCURIAL THERMOMETERS.
The mercurial thermometers used for measuring the temperature
differences of the water-current are of special construction and have
been calibrated with the greatest accuracy. As the water enters the
respiration chamber through a horizontal tube, the thermometers are so
constructed and so placed in the horizontal tubes through which the
water passes that the bulbs of the thermometers lie about in a plane
with the copper wall, thus taking the temperature of the water
immediately as it enters and as it leaves the chamber. For convenience
in reading, the stem of the thermometer is bent at right angles and the
graduations are placed on the upright part.
The thermometers are graduated from 0° to 12° C. or from 8° to 20° C.
and each degree is divided into fiftieths. Without the use of a lens it
is possible to read accurately to the hundredth of a degree. For
calibrating these thermometers a special arrangement is necessary. The
standards used consist of well-constructed metastatic thermometers of
the Beckmann type, made by C. Richter, of Berlin, and calibrated by the
Physikalische Technische Reichsanstalt. Furthermore, a standard
thermometer, graduated from 14° to 24° C., also made by Richter and
standardized by the Physikalische Technische Reichsanstalt, serves as a
basis for securing the absolute temperature. Since, however, on the
mercurial thermometers used in the water-current, differences in
temperature are required rather than absolute temperatures, it is
unnecessary, except in an approximate way, to standardize the
thermometers on the basis of absolute temperature. For calibrating the
thermometers, an ordinary wooden water-pail is provided with several
holes in the side near the bottom. One-hole rubber stoppers are inserted
in these holes and through these are placed the bulbs and stems of the
different thermometers which are to be calibrated. The upright portion
of the stem is held in a vertical position by a clamp. The pail is
filled with water, thereby insuring a large mass of water and slow
temperature fluctuations, and the water is stirred by means of an
electrically driven turbine stirrer.
The Beckmann thermometers, of which two are used, are so adjusted that
they overlap each other and thus allow a range of 8° to 14° C. without
resetting. For all temperatures above 14° C., the standard Richter
thermometer can be used directly. For temperatures at 8° C. or below, a
large funnel filled with cracked ice is placed with the stem dipping
into the water. As the ice melts, the cooling effect on the large mass
of water is sufficient to maintain the temperature constant and
compensate the heating effect of the surrounding room-air. The
thermometers are tapped and read as nearly simultaneously as possible. A
number of readings are taken at each point and the average readings used
in the calculations. Making due allowance for the corrections on the
Beckmann thermometers, the temperature differences can be determined to
less than 0.01° C. The data obtained from the calibrations are therefore
used for comparison and a table of corrections is prepared for each set
of thermometers used. It is especially important that these thermometers
be compared among themselves with great accuracy, since as used in the
calorimeter the temperature of the ingoing water is measured on one
thermometer and the temperature of the outgoing water on another.
Thermometers of this type are extremely fragile. The long angle with an
arm some 35 centimeters in length makes it difficult to handle them
without breakage, but they are extremely sensitive and accurate and have
given great satisfaction. The construction of the bulb is such, however,
that the slightest pressure on it raises the column of mercury very
perceptibly, and hence it is important in practical use to note the
influence of the pressure of the water upon the bulbs and make
corrections therefor. The influence of such pressure upon thermometers
used in an apparatus of this type was first pointed out by Armsby,[8]
and with high rates of flow, amounting to 1 liter or more per minute,
there may be a correction on these thermometers amounting to several
hundredths of a degree. We have found that, as installed at present,
with a rate of flow of less than 400 cubic centimeters per minute, there
is no correction for water pressure.
In installing a thermometer it is of the greatest importance that there
be no pressure against the side of the tube through which the
thermometer is inserted. The slightest pressure will cause considerable
rise in the mercury column. Special precautions must also be taken to
insulate the tube through which the water passes, as the passage of the
water along the tube does not insure ordinarily a thorough mixing, and
by moving the thermometer bulb from the center of the tube to a point
near the edge, the water, which at the edge may be somewhat warmer than
at the center, immediately affects the thermometer. By use of the vacuum
jacket mentioned above, this warming of the water has been avoided, and
in electric-resistance thermometers special precautions are taken not
only with regard to the relative position of the bulb of the mercury
thermometer and the resistance thermometer, but also with regard to the
hard-rubber insulation, to avoid errors of this nature.
ELECTRIC-RESISTANCE THERMOMETERS.
Electric-resistance thermometers are used in connection with the
respiration calorimeter for several purposes: first, to determine the
fluctuations in the temperature of the air inside the chamber; second,
to measure the fluctuations of the temperature of the copper wall of the
respiration chamber; third, for determining the variations in body
temperature; finally, for recording the differences in temperature of
the incoming and outgoing water. While these thermometers are all built
on the same principle, their installation is very different, and a word
regarding the method of using each is necessary.
AIR THERMOMETERS.
The air thermometers are designed with a special view to taking quickly
the temperature of the air. Five thermometers, each having a resistance
of not far from 4 ohms, are connected in series and suspended 3.5
centimeters from the wall on hooks inside the chamber. They are
surrounded for protection, first, with a perforated metal cylinder, and
outside this with a wire guard.
[Illustration: FIG. 15.--Detail of air-resistance thermometer, showing
method of mounting and wiring the thermometer. Parts of the wire guard
and brass guard are shown, cut away so that interior structure can be
seen.]
The details of construction and method of installation are shown in fig.
15. Four strips of mica are inserted into four slots in a hard maple rod
12.5 centimeters long and 12 millimeters in diameter, and around each
strip is wound 5 meters of double silk-covered pure copper wire
(wire-gage No. 30). By means of heavy connecting wires, five of these
thermometers are connected in series, giving a total resistance of the
system of not far from 20 ohms. The thermometer proper is suspended
between two hooks by rubber bands and these two hooks are in turn
fastened to a wire guard which is attached to threaded rods soldered to
the inner surface of the copper wall, thus bringing the center of the
thermometer 3.4 centimeters from the copper wall. Two of these
thermometers are placed in the dome of the calorimeter immediately over
the shoulders of the subject, and the other three are distributed around
the sides and front of the chamber. This type of construction gives
maximum sensibility to the temperature fluctuations of the air itself
and yet insures thorough protection. The two terminals are carried
outside of the respiration chamber to the observer's table, where the
temperature fluctuations are measured on a Wheatstone bridge.
WALL THERMOMETERS.
The wall thermometers are designed for the purpose of taking the
temperature of the copper wall rather than the temperature of the air.
When temperature fluctuations are being experienced inside of the
respiration chamber, the air obviously shows temperature fluctuations
first, and the copper walls are next affected. Since in making
corrections for the hydrothermal equivalent of the apparatus and for
changes in the temperature of the apparatus as a whole it is desirable
to know the temperature changes of the wall rather than the air, these
wall thermometers were installed for this special purpose. In
construction they are not unlike the thermometers used in the air, but
instead of being surrounded by perforated metal they are encased in
copper boxes soldered directly to the wall. Five such thermometers are
used in series and, though attached permanently to the wall, they are
placed in relatively the same position as the air thermometers. The two
terminals are conducted through the metal walls to the observer's table,
where variations in resistance are measured. The resistance of the five
thermometers is not far from 20 ohms.
ELECTRICAL RECTAL THERMOMETER.
The resistance thermometer used for measuring the temperature of the
body of the man is of a somewhat different type, since it is necessary
to wind the coil in a compact form, inclose it in a pure silver tube,
and connect it with suitable rubber-covered connections, so that it can
be inserted deep in the rectum. The apparatus has been described in a
number of publications.[9] The resistance of this system is also not far
from 20 ohms, thus simplifying the use of the apparatus already
installed on the observer's table.
ELECTRIC-RESISTANCE THERMOMETERS FOR THE WATER-CURRENT.
The measurement of the temperature differences of the water-current by
the electric-resistance thermometer was tried a number of years ago by
Rosa,[10] but the results were not invariably satisfactory and in all
the subsequent experimenting the resistance thermometer could not be
used with satisfaction. More recently, plans were made to incorporate
some of the results of the rapidly accumulating experience in the use of
resistance thermometers and consequently an electric-resistance
thermometer was devised to meet the conditions of experimentation with
the respiration calorimeter by Dr. E. F. Northrup, of the Leeds &
Northrup Company, of Philadelphia. The conditions to be met were that
the thermometers should take rapidly the temperature of the ingoing and
outcoming water and that the fluctuations in temperature difference as
measured by the resistance thermometers should be controlled for
calibration purposes by the differences in temperature of the mercurial
thermometers.
[Illustration: FIG. 16.--Details of resistance thermometers for
water-circuit. Upper part of figure shows a sketch of the outside of the
hard-rubber case. In lower part is a section showing interior
construction. Flattened lead tube wound about central brass tube
contains the resistance wire. A is enlarged part of the case forming a
chamber for the mercury bulb. Arrows indicate direction of flow on
resistance thermometer for ingoing water.]
For the resistance thermometer, Dr. Northrup has used, instead of
copper, pure nickel wire, which has a much higher resistance and thus
enables a much greater total resistance to be inclosed in a given space.
The insulated nickel wire is wound in a flattened spiral and then passed
through a thin lead tube flattened somewhat. This lead tube is then
wound around a central core and the flattened portions attached at such
an angle that the water passing through the tubes has a tendency to be
directed away from the center and against the outer wall, thus insuring
a mixing of the water. Space is left for the insertion of the mercurial
thermometer. With the thermometer for the ingoing water, it was found
necessary to extend the bulb somewhat beyond the resistance coil, so
that the water might be thoroughly mixed before reaching the bulb and
thus insure a steady temperature. Thus it was found necessary to enlarge
the chamber A (fig. 16) somewhat and the tube leading out of the
thermometer, so that the bulb of the thermometer itself could be placed
almost directly at the opening of the exit tube. Under these conditions
perfect mixing of water and constancy of temperature were obtained.
In the case of the thermometer which measured the outcoming water, the
difficulty was not so great, as the outcoming water is somewhat nearer
the temperature of the chamber, and the water as it leaves the
thermometer passes first over the mercurial thermometer and then over
the resistance thermometer. By means of a long series of tests it was
found possible to adjust these resistance thermometers so that the
variations in resistance were in direct proportion to the temperature
changes noted on the mercurial thermometers. Obviously, these
differences in resistance of the two thermometers can be measured
directly with the Wheatstone bridge, but, what is more satisfactory,
they are measured and recorded directly on a special type of automatic
recorder described beyond.
OBSERVER'S TABLE.
The measurements of the temperature of the respiration chamber, of the
water-current, and of the body temperature of the man, as well as the
heating and cooling of the air-spaces about the calorimeter, are all
under the control of the physical assistant. The apparatus for these
temperature controls and measurements is all collected compactly on a
table, the so-called "observer's table." At this, the physical assistant
sits throughout the experiments. For convenience in observing the
mercurial thermometers in the water-current and general inspection of
the whole apparatus, this table is placed on an elevated platform, shown
in fig. 3. Directly in front of the table the galvanometer is suspended
from the ceiling and a black hood extends from the observer's table to
the galvanometer itself. On the observer's table proper are all the
electrical connections and at the left are the mercurial thermometers
for the chair calorimeter. Formerly, when the method of alternately
cooling and heating the air-spaces was used, the observer was able to
open and close the water-valves without leaving the chair.
The observer's table is so arranged electrically as to make possible
temperature control and measurement of either of the two calorimeters.
It is impossible, however, for the observer to read the mercurial
thermometers in the bed calorimeter without leaving his chair, and
likewise he must occasionally alter the cooling water flowing through
the outer air-spaces by going to the bed calorimeter itself. The
installation of the electric-resistance thermometers connected with the
temperature recorder does away with the reading of the mercurial
thermometers, save for purposes of comparison, and hence it is
unnecessary for the assistant to leave the chair at the observer's table
when the bed calorimeter is in use. Likewise the substitution of the
method of continuously cooling somewhat the air-spaces and reheating
with electricity, mentioned on page 18, does away with the necessity for
alternately opening and closing the water-valves of the chair
calorimeter placed at the left of the observer's table.
[Illustration: FIG. 17.--Diagram of wiring of observer's table. W_{1},
W_{2}, Wheatstone bridges for resistance thermometers; K_{1}, K_{2},
double contact keys for controlling Wheatstone circuits; S_{1}, S_{2},
S_{3}, double-pole double-throw switches for changing from chair to bed
calorimeter; S_{4}, double-pole double-throw switch for changing from
wall to air thermometers; G, galvanometer; R_{2}, rheostat. 1, 2, 3, 4,
5, wires connecting with resistance-coils A B D E F and _a b d e f_;
S_{2}, 6-point switch for connecting thermal-junction circuits of either
bed or chair calorimeter with galvanometer; S_{10}, 10-point
double-throw switch for changing heating circuits and thermal-junction
circuits to either chair or bed calorimeter; R_{1}, rheostat for
controlling electric heaters in ingoing water in calorimeters; S_{8},
double-pole single-throw switch for connecting 110-v. current with
connections on table; S_{9}, double-pole single-throw switch for
connecting R_{1} with bed calorimeter.]
Of special interest are the electrical connections on the observer's
table itself. A diagrammatic representation of the observer's table with
its connections is shown in fig. 17. The heavy black outline gives in a
general way the outline of the table proper and thus shows a
diagrammatic distribution of the parts. The first of the electrical
measurements necessary during experiments is that of the
thermo-electric effect of the thermal junction systems installed on the
calorimeters. To aid in indicating what parts of the zinc wall need
cooling or heating, the thermal junction systems are, as has already
been described, separated into four sections on the chair calorimeter
and three sections on the bed calorimeter; in the first calorimeter, the
top, front, rear, and bottom; in the bed calorimeter, the top, sides,
and bottom.
CONNECTIONS TO THERMAL-JUNCTION SYSTEMS.
Since heretofore it has been deemed unwise to attempt to use both
calorimeters at the same time, the electrical connections are so made
that, by means of electrical switches, either calorimeter can be
connected to the apparatus on the table.
The thermal-junction measurements are made by a semicircular switch
S_{7}. The various points, I, II, III, IV, etc., are connected with the
different thermal-junction systems. Thus, by following the wiring
diagram, it can be seen that the connections with I run to the different
binding-posts of the switch S_{10}, which as a matter of fact is placed
beneath the table. This switch S_{10} has three rows of binding-posts.
The center row connects directly with the apparatus on the observer's
table, the outer rows connect with either the chair calorimeter or the
bed calorimeter. The points marked _a_, _b_, _d_, _e_, _f_, etc.,
connect with the bed calorimeter and A, B, D, etc., connect with the
chair calorimeter. Thus, by connecting the points _g_ and _i_ with the
two binding-posts opposite them on the switch S_{10}, it can be seen
that this connection leads directly to the point I on the switch S_{7},
and as a matter of fact this gives direct connection with the
galvanometer through the key on S_{7}, thus connecting the
thermal-junction system on one section of the bed calorimeter between
_g_ and _i_ directly with the galvanometer. Similar connections from the
other points can readily be followed from the diagram. The points on the
switch S_{7} indicated as I, II, III, IV, correspond respectively to the
thermal-junction systems on the top, rear, front, and bottom of the
chair calorimeter.
By following the wiring diagram of the point V, it will be seen that
this will include the connections with the thermal junctions connected
in series and thus give a sum total of the electromotive forces in the
thermal junctions. The point VI is connected with the thermal-junction
system in the air system, indicating the differences in temperature
between the ingoing and outgoing air. It will be noted that there are
four sections in the chair calorimeter, while in the bed calorimeter
there are but three, and hence a special switch S_{3} is installed to
insure proper connections when the bed calorimeter is in use.
This system of connecting the thermal junctions in different sections to
the galvanometer makes possible a more accurate control of the
temperatures in the various parts, and while the algebraic sum of the
temperature differences of the parts may equal zero, it is conceivable
that there may be a condition in the calorimeter when there is a
considerable amount of heat passing out through the top, for example,
compensated exactly by the heat which passes in at the bottom, and while
with the top section there would be a large plus deflection on the
galvanometer, thus indicating that the air around the zinc wall was too
cold and that heat was passing out, there would be a corresponding minus
deflection on the bottom section, indicating the reverse conditions. The
two may exactly balance each other, but it has been found advantageous
to consider each section as a unit by itself and to attempt delicate
temperature control of each individual unit. This has been made possible
by the electrical connections, as shown on the diagram.
RHEOSTAT FOR HEATING.
The rheostat for heating the air-spaces and the returning air-current
about the zinc wall is placed on the observer's table and is indicated
in the diagram as R_{2}. There are five different sets of
contact-points, marked 1, 2, 3, 4, and 5. One end of the rheostat is
connected directly with the 110-volt circuit through the main switch
S_{5}. The other side of the switch S_{5} connects directly with the
point on the middle of switch S_{10}, and when this middle point is
joined with either _f_ and F, direct connection is insured between all
the various heating-circuits on the calorimeter in use. The various
numbered points on the rheostat R_{2}, are connected with the binding
posts on S_{10}, and each can in turn be connected with _a_ or A, _b_ or
B, etc. The heating of the top of the chair calorimeter is controlled by
the point 5 on the rheostat R_{2}, the rear by the point 4, the front by
the point 3, and the bottom by the point 2. Point 1 is used for heating
the air entering the calorimeter by means of an electric lamp placed in
the air-pipe, as shown in fig. 25.
The warming of the electrical reheater placed in the water-circuit just
before the water enters the calorimeter is done by an electrical current
controlled by the resistance R_{1}. This R_{1} is connected on one end
directly with the 110-volt circuit and the current leaving it passes
through the resistance inside the heater in the water-current. The two
heaters, one for each calorimeter, are indicated on the diagram above
and below the switch S_{9}. The disposition of the switches is such as
to make it possible to use alternately the reheaters on either the bed
or the chair calorimeter, and the main resistance R_{1} suffices for
both.
WHEATSTONE BRIDGES.
For use in measuring the temperature of the air and of the copper wall
of the calorimeters, as well as the rectal temperature of the subject, a
series of resistance thermometers is employed. These are so connected
on the observer's table that they may be brought into connection with
two Wheatstone bridges, W_{1} and W_{2}. Bridge W_{1} is used for the
resistance thermometers indicating the temperature of the wall and the
air. Bridge W_{2} is for the rectal thermometer. Since similar
thermometers are inserted in both calorimeters, it is necessary to
introduce some switch to connect either set at will and hence the
double-throw switches S_{1}, S_{2}, and S_{3} allow the use of either
the wall, air, or rectal thermometer on either the bed or chair
calorimeter at will. Since the bridge W_{1} is used for measuring the
temperature of both the wall and the air, a fourth double-pole switch,
S_{4}, is used to connect the air and wall thermometers alternately. The
double-contact key, K_{1}, is connected with the bridge W_{1} and is so
arranged that the battery circuit is first made and subsequently the
galvanometer circuit. A similar arrangement in K_{2} controls the
connections for the bridge W_{2}.
GALVANOMETER.
The galvanometer is of the Deprez-d'Arsonval type and is extremely
sensitive. The sensitiveness is so great that it is desirable to
introduce a resistance of some 500 ohms into the thermal-junction
circuits. This is indicated at the top of the diagram near the
galvanometer. The maximum sensitiveness of the galvanometer is retained
when the connection is made with the Wheatstone bridges. The
galvanometer is suspended from the ceiling of the calorimeter laboratory
and is free from vibration.
RESISTANCE FOR HEATING COILS.
To vary the current passing through the manganin heating coils in the
air-spaces next the zinc wall, a series of resistances is installed
connected directly with the rheostat R_{2} in fig. 17. The details of
these resistances and their connection with the rheostat are shown in
fig. 18. The rheostat, which is in the right part of the figure, has
five sliding contacts, each of which can be connected with ten different
points. One end of the rheostat is connected directly with the 110-volt
circuit. Beneath the observer's table are fastened the five resistances,
which consist of four lamps, each having approximately 200 ohms
resistance and then a series of resistance-coils wound on a long strip
of asbestos lumber, each section having approximately 15 ohms between
the binding-posts. A fuse-wire is inserted in each circuit to protect
the chamber from excessive current. Of these resistances, No. 1 is used
to heat the lamp in the air-current shown in fig. 25, and consequently
it has been found advisable to place permanently a second lamp in series
with the first, but outside of the air-pipe, so as to avoid burning out
the lamp inside of the air-pipe. The other four resistances, 2, 3, 4,
and 5, are connected with the different sections on the two
calorimeters. No. 5 corresponds to the top of both calorimeters. No. 4
corresponds to the rear section of the chair calorimeter and to the
sides of the bed calorimeter. No. 3 corresponds to the front of the
chair calorimeter and is without communication with the bed calorimeter.
No. 2 connects with the bottom of both calorimeters.
It will be seen from the diagrams that each of these resistances can be
connected at will with either the bed or the chair calorimeter and at
such points as are indicated by the lettering below the numbers. Thus,
section 1 can be connected with either the point A or point _a_ on fig.
17 and thus directly control the amount of current passing through the
corresponding resistance in series with the lamp in the air-current. The
sliding contacts at present in use are ill adapted to long-continued
usage and will therefore shortly be substituted by a more substantial
instrument. The form of resistance using small lamps and the resistance
wires wound on asbestos lumber has proven very satisfactory and very
compact in form.
[Illustration: FIG. 18.--Diagram of rheostat and resistances in series
with it. At the right are shown the sliding contacts, and in the center
places for lamps used as resistances, and to left the sections of wire
resistances.]
TEMPERATURE RECORDER.
The numerous electrical, thermometric, and chemical measurements
necessary in the full conduct of an experiment with the respiration
calorimeter has often raised the question of the desirability of making
at least a portion of these observations more or less automatic. This
seems particularly feasible with the observations ordinarily recorded by
the physical observer. These observations consist of the reading of the
mercurial thermometers indicating the temperatures of the ingoing and
outcoming water, records with the electric-resistance thermometers for
the temperature of the air and the walls and the body temperatures, and
the deflections of the thermo-electric elements.
Numerous plans have been proposed for rendering automatic some of these
observations, as well as the control of the heating and cooling of the
air-circuits. Obviously, such a record of temperature measurements would
have two distinct advantages: (1) in giving an accurate graphic record
which would be permanent and in which the influence of the personal
equation would be eliminated; (2) while the physical observer at present
has much less to do than with the earlier form of apparatus, it would
materially lighten his labors and thereby tend to minimize errors in the
other observations.
The development of the thread recorder and the photographic registration
apparatus in recent years led to the belief that we could employ similar
apparatus in connection with our investigations in this laboratory. To
this end a number of accurate electrical measuring instruments were
purchased, and after a number of tests it was considered feasible to
record automatically the temperature differences of the ingoing and
outcoming water from the calorimeter. Based upon our preliminary tests,
the Leeds & Northrup Company of Philadelphia, whose experience with such
problems is very extended, were commissioned to construct an apparatus
to meet the requirements of the respiration calorimeter. The conditions
to be met by this apparatus were such as to call for a registering
recorder that would indicate the differences in temperature between the
ingoing and outcoming water to within 0.5 per cent and to record these
differences in a permanent ink line on coordinate paper. Furthermore,
the apparatus must be installed in a fixed position in the laboratory,
and connections should be such as to make it interchangeable with any
one of five calorimeters.
After a great deal of preliminary experimenting, in which the Leeds &
Northrup Company have most generously interpreted our specifications,
they have furnished us with an apparatus which meets to a high degree of
satisfaction the conditions imposed. The thermometers themselves have
already been discussed. (See page 30.) The recording apparatus consists
of three parts: (1) the galvanometer; (2) the creeper or automatic
sliding-contact; (3) the clockwork for the forward movement of the roll
of coordinate paper and to control the periodic movement of the creeper.
Under ordinary conditions with rest experiments in the chair calorimeter
or bed calorimeter, the temperature differences run not far from 2° to
4°. Thus, it is seen that if the apparatus is to meet the conditions of
the specifications it must measure differences of 2° C. to within 0.01°
C. Provision has also been made to extend the measurement of temperature
differences with the apparatus so that a difference of 8° can be
measured with the same percentage accuracy.
FUNDAMENTAL PRINCIPLE OF THE APPARATUS.
The apparatus depends fundamentally upon the perfect balancing of the
two sides of a differential electric circuit. A conventional diagram,
fig. 19, gives a schematic outline of the connections. The two
galvanometer coils, _fl_ and _fr_, are wound differentially and both
coils most carefully balanced so that the two windings have equal
temperature coefficients. This is done by inserting a small shunt _y_,
parallel with the coil _fl_, and thus the temperature coefficient of
_fl_ and _fr_ are made absolutely equal. The two thermometers are
indicated as T_{1} and T_{2} and are inserted in the ingoing and
outgoing water respectively. A slide-wire resistance is indicated by J,
and _r_ is the resistance for the zero adjustment. Ba, Z, and Z_{1} are
the battery and its variable series resistances. If T_{1} and T_{2} are
exactly of the same temperature, _i. e._, if the temperature difference
of the ingoing and outcoming water is zero, the sliding contact _q_
stands at 0 on the slide-wire and thus the resistance of the system from
0 through _fl_, _r_, and T_{1} back to the point C is exactly the same
as the resistance of the slide-wire J plus the coil _fr_ plus T_{2} back
to the point C. A rise in temperature of T_{2} gives an increase of
resistance in the circuit and the sliding contact _q_ moves along the
slide-wire toward J maximum until a balance is obtained.
[Illustration: FIG. 19.--Diagram of wiring of differential circuit with
its various shunts, used in connection with resistance thermometers on
water-circuit of bed calorimeter.]
Provision is made for automatically moving the contact _q_ by electrical
means and thus the complete balance of the two differential circuits is
maintained constant from second to second. As the contact _q_ is moved,
it carries with it a stylographic pen which travels in a straight line
over a regularly moving roll of coordinate paper, thus producing a
permanently recorded curve indicating the temperature differences. The
slide-wire J is calibrated so that any inequalities in the temperature
coefficient of the thermometer wires are equalized and also so that any
unit-length on the slide-wire taken at any point along the temperature
scale represents a resistance equal to the resistance change in the
thermometer for that particular change in temperature. With the varying
conditions to be met with in this apparatus, it is necessary that
varying values should be assigned at times to J and to _r_. This
necessitates the use of shunts, and the recording range of the
instrument can be easily varied by simple shunting, _i. e._, by changing
the resistance value of J and _r_, providing these resistances unshunted
have a value which takes care of the highest obtained temperature
variations.
Fig. 19 shows the differential circuit complete with all its shunts. S
is a fixed shunt to obtain a range on J; S' is a variable shunt to
permit very slight variations of J within the range to correct errors
due to changing of the initial temperatures of the thermometers; _y_ is
a permanent shunt across the galvanometer coil _fl_, to make the
temperature coefficients of _fl_ and _fr_ absolutely equal; Z is the
variable resistance in the battery-circuit to keep the current constant;
_r_ is a permanent resistance to fix the zero on varying ranges; S''
plus S_{1} constitutes a variable shunt to permit slight variations of
_r_ to finally adjust 0 after S' is fixed and _t_ is a permanent shunt
across the thermometer T_{1} to make the temperature coefficient of
T_{1} equal to that of T_{2}.
The apparatus can be used for measuring temperature differences from 0°
to 4° or from 0° to 8°. When on the 0° to 8° range, the shunt S is
open-circuited and the shunt S' alone used. The value of S, then, is
predetermined so as to affect the value of the wire J and thus halve its
influence in maintaining the balance. Similarly, when the lower range,
_i. e._, from 0° to 4°, is used, the resistance _r_ is employed, and
when the higher range is used another value to _r_ must be given by
using a plug resistance-box, in the use of which the resistance _r_ is
doubled.
The resistance S'' and S_{1} are combined in a slide-wire resistance-box
and are used to change the value of the whole apparatus when there are
marked changes in the position of the thermometric scale. Thus, if the
ingoing water is at 2° C. and the outcoming water at 5° C. in one
instance, and in another instance the ingoing water is 13° and the
outgoing water is 15°, a slight alteration in the value of S_{1}, and
also of S', is necessary in order to have the apparatus draw a curve to
represent truly the temperature differences. These slight alterations
are determined beforehand by careful tests and the exact value of the
resistances in S' and in S_{1} are permanently recorded for subsequent
use.
THE GALVANOMETER.
The galvanometer is of the Deprez-d'Arsonval type and has a particularly
powerful magnetic field, in which a double coil swings suspended similar
to the marine galvanometer coils. This coil is protected from vibrations
by an anti-vibration tube A, fig. 20, and carries a pointer P which acts
to select the direction of movement of the recording apparatus, the
movable contact point _q_, fig. 19. In front of this galvanometer coil
and inclosed in the same air-tight metal case is the plunger contact Pl,
fig. 21. The galvanometer pointer P swings freely below the silver
contacts S_{1} and S_{2}, just clearing the ivory insulator _i_. The
magnet plunger makes a contact depending upon the adjustment of a clock
at intervals of 2 seconds. So long as both galvanometer coils are
influenced by exactly the same strength of current, the pointer will
stand in line with and immediately below _i_ and no current passes
through the recording apparatus. Any disturbance of the electrical
equilibrium causes the pointer P to swing either toward S_{1} or S_{2},
thus completing the circuit at either the right hand or the left hand,
at intervals of 2 seconds. The movement of the pointer away from its
normal position exactly beneath _i_ to either S_{1} on the left hand or
S_{2} on the right, results from an inequality in the current flowing
through the two coils in the galvanometer. The difference in the two
currents passing through these coils is caused by a change in
temperatures of the two thermometers in the water circuit.
[Illustration: FIG. 20.--Diagram of galvanometer coil used in connection
with recording apparatus for resistance thermometers in the
water-circuit of bed calorimeter. A, anti-vibration tube; P, pointer.]
THE CREEPER.
The movement of the sliding-contact _q_, fig. 19, along the slide-wire
J, is produced by means of a special device called a creeper, consisting
of a piece of brass carefully fitted to a threaded steel rod some 30
centimeters long. The movement of this bar along this threaded rod
accomplishes two things. The bar is in contact with the slide-wire J
and therefore varies the position of the point _q_ and it also carries
with it a stylographic pen. The movements of this bar to the right or
the left are produced by an auxiliary electric current, the contact of
which is made by a plunger-plate forcing the pointer P against either
S_{1} or S_{2}. P makes the contact between Pl and either S_{1} or S_{2}
and sends a current through solenoids at either the right or the left of
the creeper. At intervals of every 2 seconds the plunger rises and
forces the pointer P against either S_{1}, _i_, or S_{2} above. The
movement of this plunger is controlled by a current from a 110-volt
circuit, the connections of which are shown in fig. 22. If the contact
is made at T, the current passes through 2,600 ohms, directly across the
110-volt circuit, and consequently there is no effective current flowing
through the plunger Pl. When the contact T is open, the current flows
through the plunger in series with 2,600 ohms resistance. T is opened
automatically at intervals of 2 seconds by the clock.
[Illustration: FIG. 21.--Diagram of wiring of circuits actuating plunger
and creeper.]
[Illustration: FIG. 22.--Diagram of wiring of complete 110-volt
circuit.]
The movement of the contact arm along the threaded rod is produced by
the action of either one of two solenoids, each of which has a core
attached to a rack and pinion at either end of the rod. If the current
is passed through the contact S_{1}, a current passes through the
left-hand solenoid, the core moves down, the rack on the core moves the
pinion on the rod through a definite fraction of a complete revolution
and this movement forces the creeper in one direction. Conversely, the
passing of the current through the solenoid at the other end of the
threaded rod moves the creeper in the other direction. The distance
which the iron rack on the end of the core is moved is determined
carefully, so that the threaded rod is turned for each contact exactly
the same fraction of a revolution. For actuating these solenoids, the
110-volt circuit is again used. The wire connections are shown in part
in fig. 21, in which it is seen that the current passes through the
plunger-contact and through the pointer P to the silver plate S_{1} and
then along the line G_{1} through 350 ohms wound about the left-hand
solenoid back through a 600-ohm resistance to the main line. The use of
the 110-volt current under such circumstances would normally produce a
notable sparking effect on the pointer P, and to reduce this to a
minimum there is a high resistance, amounting to 10,000 ohms on each
side, shunted between the main line and the creeper connections. This
shunt is shown in diagram in fig. 22. Thus there is never a complete
open circuit and sparking is prevented.
THE CLOCK.
The clock requires winding every week and is so geared as to move the
paper forward at a rate of 3 inches per hour. The contact-point for
opening the circuit T on fig. 22 is likewise connected with one of the
smaller wheels of the clock. This contact is made by tripping a little
lever by means of a toothed wheel of phosphor-bronze.
INSTALLATION OF THE APPARATUS.
[Illustration: FIG. 23
Temperature recorder. The recorder with the coordinate paper in the
lower box with a glass door. A curve representing the temperature
difference between the ingoing and outgoing water is directly drawn on
the coordinate paper. Above are three resistance boxes, and the switches
for electrical connections are at the right. On the top shelf is the
galvanometer, and immediately beneath, the plug resistance box for
altering the value of certain shunts.]
[Illustration: FIG. 24.--Detailed wiring diagram showing all parts of
recording apparatus, together with wiring to thermometers complete,
including all previous figures.]
The whole apparatus is permanently and substantially installed on the
north wall of the calorimeter laboratory. A photograph showing the
various parts and their installation is given in fig. 23. On the top
shelf is seen the galvanometer and on the lower shelf the recorder with
its glass door in front and the coordinate paper dropping into the box
below. The curve drawn on the coordinate paper is clearly shown. Above
the recorder are the resistance-boxes, three in number, the lower one at
the left being the resistance S_{1}, the upper one at the left being the
resistance S', and the upper one at the right being the resistance
Z_{1}. Immediately above the resistance-box Z_{1} is shown the plug
resistance-box which controls on the one hand the resistance _r_ and on
the other hand the resistance S, both of which are substantially altered
when changing the apparatus to register from the 0° to 4° scale to the
0° to 8° scale. A detailed wiring diagram is given in fig. 24.
TEMPERATURE CONTROL OF THE INGOING AIR.
[Illustration: FIG. 25.--Section of calorimeter walls and part of
ventilating air-circuit, showing part of pipes for ingoing air and
outgoing air. On the ingoing air-pipe at the right is the lamp for
heating the ingoing air. Just above it, H is the quick-throw valve for
shutting off the tension equalizer IJ. I is the copper portion of the
tension equalizer, while J is the rubber diaphragm; K, the pet-cock for
admitting oxygen; F, E, G, the lead pipe conducting the cold water for
the ingoing air; and C, the hair-felt insulation. N, N are brass ferules
soldered into the copper and zinc walls through which air-pipes pass; M,
a rubber stopper for insulating the air-pipe from the calorimeter; O,
the thermal junctions for indicating differences of temperature of
ingoing and outgoing air and U, the connection to the outside; QQ, exits
for the air-pipes from the box in which thermal junctions are placed; P,
the dividing plate separating the ingoing and outgoing air; R, the
section of piping conducting the air inside the calorimeter; S, a
section of piping through which the air passes from the calorimeter; A,
a section of the copper wall; Y, a bolt fastening the copper wall to the
2-1/2 inch angle W; B, a portion of zinc wall; C, hair-felt lining of
asbestos wall D; T-J, a thermal junction in the walls.]
In passing the current of air through the calorimeter, temperature
conditions may easily be such that the air entering is warmer than the
outcoming air, in which case heat will be imparted to the calorimeter,
or the reverse conditions may obtain and then heat will be brought away.
To avoid this difficulty, arrangements are made for arbitrarily
controlling the temperature of the air as it enters the calorimeter.
This temperature control is based upon the fact that the air leaving the
chamber is caused to pass over the ends of a series of thermal junctions
shown as O in fig. 25. These thermal junctions have one terminal in the
outgoing air and the other in the ingoing air, and consequently any
difference in the temperature of the two air-currents is instantly
detected by connecting the circuit with the galvanometer. Formerly the
temperature control was made a varying one, by providing for either
cooling or heating the ingoing air as the situation called for. The
heating was done by passing the current through an electric lamp placed
in the cross immediately below the tension equalizer J. Cooling was
effected by means of a current of water through the lead pipe E closely
wrapped around the air-pipe, water entering at F and leaving at G. This
lead pipe is insulated by hair-felt pipe-covering, C. More recently, we
have adopted the procedure of passing a continuous current of water,
usually at a very slow rate, through the lead pipe E and always heating
the air somewhat by means of the lamp, the exact temperature control
being obtained by varying the heating effect of the lamp itself. This
has been found much more satisfactory than by alternating from the
cooling system to the heating system. In the case of the air-current,
however, it is unnecessary to have the drop-sight feed-valve as used for
the wall control, shown in fig. 13.
THE HEAT OF VAPORIZATION OF WATER.
During experiments with man not all the heat leaves the body by
radiation and conduction, since a part is required to vaporize the water
from the skin and lungs. An accurate measurement of the heat production
by man therefore required a knowledge of the amount of heat thus
vaporized. One of the great difficulties in the numerous forms of
calorimeters that have been used heretofore with man is that only that
portion of heat measured by direct radiation or conduction has been
measured and the difficulties attending the determination of water
vaporized have vitiated correspondingly the estimates of the heat
production. Fortunately, with this apparatus the determinations of water
are very exact, and since the amount of water vaporized inside the
chamber is known it is possible to compute the heat required to vaporize
this water by knowing the heat of vaporization of water.
Since the earlier reports describing the first form of calorimeters were
written, there has appeared a research by one of our former associates,
Dr. A. W. Smith[11] who, recognizing the importance of knowing exactly
the heat of vaporization of water at 20°, has made this a special object
of investigation. When connected with our laboratory a number of
experiments were made by Doctors Smith and Benedict in an attempt to
determine the heat of vaporization of water directly in a large
calorimeter; but for lack of time and pressure of other experimental
work it was impossible to complete the investigation. Subsequently Dr.
Smith has carried out the experiments with the accuracy of exact
physical measurements and has given us a very valuable series of
observations.
Using the method of expressing the heat of vaporization in electrical
units, Smith concludes that the heat of vaporization of water between
14° and 40° is given by the formula
L (in joules) = 2502.5 - 2.43T
and states that the "probable error" of values computed from this
formula is 0.5 joule. The results are expressed in international joules,
that is, in terms of the international ohm and 1.43400 for the E.M.F. of
the Clark cell at 15° C., and assuming that the mean calorie is
equivalent to 4.1877 international joules,[12] the formula reads
L (in mean calories) = 597.44 - 0.580T
With this formula Smith calculates that at 15° the heat of vapor
Cite this
Francis Gano Benedict. "Respiration Calorimeters for Studying the Respiratory Exchange and Energy Transformations of Man". Project Gutenberg. Public domain in the USA. https://www.gutenberg.org/ebooks/28216 JSON record · API collection · BibTeX · RIS · CSL-JSON · citation policy