Pyrometry: A Practical Treatise on the Measurement of High Temperatures
315,452 chars · English
The Project Gutenberg eBook of Pyrometry: A Practical Treatise on the Measurement of High Temperatures
This eBook is for the use of anyone anywhere in the United States and
most other parts of the world at no cost and with almost no restrictions
whatsoever. You may copy it, give it away or re-use it under the terms
of the Project Gutenberg License included with this eBook or online
at www.gutenberg.org. If you are not located in the United States,
you will have to check the laws of the country where you are located
before using this eBook.
Title: Pyrometry: A Practical Treatise on the Measurement of High Temperatures
Author: Charles R. Darling
Release date: January 10, 2018 [eBook #56353]
Language: English
Other information and formats: www.gutenberg.org/ebooks/56353
Credits: Produced by Chris Curnow, Paul Marshall and the Online
Distributed Proofreading Team at http://www.pgdp.net (This
file was produced from images generously made available
by The Internet Archive
*** START OF THE PROJECT GUTENBERG EBOOK PYROMETRY: A PRACTICAL TREATISE ON THE MEASUREMENT OF HIGH TEMPERATURES ***
Produced by Chris Curnow, Paul Marshall and the Online
Distributed Proofreading Team at http://www.pgdp.net (This
file was produced from images generously made available
by The Internet Archive
Transcriber’s Notes:
Underscores “_” before and after a word or phrase indicate _italics_
in the original text.
Carat symbol “^” designates a superscript.
A single underscore between two characters indicates a subscript.
Equal signs “=” before and after a word or phrase indicate =bold=
in the original text.
Small capitals have been converted to SOLID capitals.
Illustrations have been moved so they do not break up paragraphs.
Typographical errors have been silently corrected but other variations
in spelling and punctuation remain unaltered.
PYROMETRY
_BY THE SAME AUTHOR_
HEAT FOR ENGINEERS
A TREATISE ON HEAT WITH
SPECIAL REGARD TO ITS
PRACTICAL APPLICATIONS
Third Edition, revised, with 110 illustrations,
xiv + 430 pp. Demy 8vo. Price 12s. 6d. net.
LIQUID DROPS AND
GLOBULES
THEIR FORMATION AND MOVEMENT
THREE LECTURES DELIVERED
TO POPULAR AUDIENCES
With 43 illustrations, x + 84 pp. Crown 8vo,
cloth. Price 3s. net.
E. & F. N. SPON, LTD., HAYMARKET, LONDON, S.W. 1
PYROMETRY
A PRACTICAL TREATISE ON
THE MEASUREMENT OF HIGH
TEMPERATURES
BY
CHAS. R. DARLING
ASSOCIATE OF THE ROYAL COLLEGE OF SCIENCE, DUBLIN; WHITWORTH
EXHIBITIONER; FELLOW OF THE INSTITUTE OF CHEMISTRY; FELLOW
OF THE PHYSICAL SOCIETY, ETC.
LECTURER IN PHYSICS AT THE CITY AND GUILDS TECHNICAL COLLEGE,
FINSBURY, E.C.
AUTHOR OF “HEAT FOR ENGINEERS”
SECOND EDITION, REVISED AND ENLARGED
SIXTY-NINE ILLUSTRATIONS
[Illustration]
_London_
E. & F. N. SPON, Ltd., 57 HAYMARKET, S.W. 1
_New York_
SPON & CHAMBERLAIN, 120 LIBERTY STREET
1920
Contents
PAGE
PREFACE TO SECOND EDITION ix
PREFACE TO FIRST EDITION xi
CHAP.
I. INTRODUCTION 1
II. STANDARDS OF TEMPERATURE 9
Absolute or Thermodynamic Scale—Constant Volume
Gas Thermometer—Fixed Points for Calibration—National
Physical Laboratory Scale—Temperatures
above the Present Limit of the Gas Thermometer.
III. THERMO-ELECTRIC PYROMETERS 20
General Principles—Metals used for Thermal Junctions—
Changes in Thermal Junctions when constantly
used—Electromotive Force developed by
Typical Junctions—Practical Forms of Thermocouples—
Liquid Element Thermocouples—Indicators for
Thermo-electric Pyrometers—Special Features of
Indicators—Standardizingof Indicators to read
Temperatures directly—Standardization by Fixed Points—
Standardization by Measurement of E.M.F.—Cold-Junction
Compensators—Constant Temperature Cold Junctions—
Special-Range Indicators—PotentiometerIndicators—
Recorders for Thermo-electric Pyrometers—The Thread
Recorder—The Siemens Recorder—Foster’s Recorder—
Paul’s Recorder—The Leeds-Northrup Recorder—
Control of Furnace Temperatures—Contact-Pen Recorders—
Installations of Thermo-electric Pyrometers—Management
of Thermo-electric Pyrometers—Laboratory Uses of
Temperatures Thermo-electric Pyrometers—Measurement of
Lower by the Thermo-electric Method—Measurement of Surface
of Surface Temperatures—Measurement of Low Temperatures—
Temperature of Steam, Exhaust Gases—Measurement of
Differences of Temperature—Advantages of the
Thermo-electric Method of Measuring Temperatures.
IV. RESISTANCE PYROMETERS 101
General Principles—Measurement of Resistance by
the Differential Galvanometer—Measurement of Resistance
by the Wheatstone Bridge—Relation between
Resistance of Platinum and Temperature—Changes
in Resistance of Platinum when constantly Heated—Terms
used in Resistance Pyrometry—Practical Forms
of Resistance Pyrometers—Indicators—Siemens’
Indicator—Whipple’s Indicator—The Harris Indicator—
The Leeds-Northrup Indicator—Siemens’ Differential
Indicator—Recorders for Resistance Pyrometers—The
Leeds-Northrup Recorder—Paul’s Recorder—Installation
of Resistance Pyrometers—Management of Resistance
Pyrometers—Special Uses of Resistance Pyrometers.
V. RADIATION PYROMETERS 134
General Principles—Practical Forms of Radiation
Pyrometers—Féry’s Mirror Pyrometer—Féry’s Spiral
Radiation Pyrometer—Foster’s Fixed-Focus Radiation
Pyrometer—Paul’s Radiation Pyrometer—Indicators
for Radiation Pyrometers—Calibration of Indicators—
Recorders—Management of Radiation Pyrometers—Special
Uses of Radiation Pyrometers.
VI. OPTICAL PYROMETERS 167
General Principles—Wien’s Law—Practical Forms of
Optical Pyrometers—Féry’s Optical Pyrometer—Le
Chatelier’s Optical Pyrometer—Wanner’s Pyrometer—
Cambridge Optical Pyrometer—Holborn-Kurlbaum
Pyrometer—Lovibond’s Pyrometer—Mesuré and
Nouel’s Pyrometer—Colour-extinction Pyrometers—
Management of Optical Pyrometers—Special Uses of
Optical Pyrometers.
VII. CALORIMETRIC PYROMETERS 195
General Principles—Practical Forms—Siemens’
Calorimetric or “Water” Pyrometer—Special Uses.
VIII. FUSION PYROMETERS 204
General Principles—Seger Pyramids or “Cones”—
Watkin’s Heat Recorder—“Sentinel” Pyrometers—
Stone’s Pyrometer—Fusible Metals—Fusible Pastes.
IX. MISCELLANEOUS APPLIANCES 211
Expansion and Contraction Pyrometers—Wedgwood’s
Pyrometer—Daniell’s Pyrometer—Northrup’s Molten
Tin Pyrometer—Vapour-Pressure Pyrometers—Water-Jet
Pyrometers—Pneumatic Pyrometers—Conduction
Pyrometers—Gas Pyrometers—Wiborgh’s Thermophones—
Joly’s Meldometer—Brearley’s Curve Tracer.
INDEX 222
Preface to the Second Edition
Since the publication of the first edition in 1911, a great extension
has been witnessed in the use of pyrometers in industrial processes
and laboratory work, to which development the author hopes his book
has contributed in some measure. During the stress occasioned by the
war, pyrometers have proved invaluable in many processes, and British
makers were fully able to meet the demands, owing to the status
attained in pre-war days. The increasing use of pyrometric appliances
renders necessary some book of reference which will provide the user
with information to enable him to get the best results out of his
instruments, and it is hoped that the present treatise meets this need.
In preparing the second edition, certain parts have been revised in
conformity with modern practice, and the later developments included.
The scope of the book remains as before.
The author desires to acknowledge the assistance he has received from
the British makers of pyrometers, all of whom have liberally provided
him with information of a most useful kind, of which he has availed
himself in the production of the present edition.
CHAS. R. DARLING.
WOOLWICH, 1920.
Preface to the First Edition
The present treatise has been founded on a course of Cantor Lectures
on “Industrial Pyrometry,” delivered by the author before the Royal
Society of Arts in the autumn of 1910. The practice of pyrometry in
recent years has progressed at a greater rate than the literature
bearing upon it; and the author is not aware of the existence of
any other book written in English which treats the subject from
the standpoint of the actual daily use of the instruments. In the
succeeding pages the exact measurement of temperature, as an end in
itself, is made subordinate to the practical utility of pyrometers
in controlling various operations; and consequently descriptions of
appliances of interest only theoretically have either been omitted, or
at the most briefly described. Nevertheless, the fundamental principles
are in all cases fully explained, as an understanding of these is
essential to the intelligent use of the appliances dealt with in the
book. When necessary, numerical examples are given to illustrate the
applications of the principles; and the reader who finds any difficulty
in following the various explanations—which of necessity involve an
understanding of many portions of the subject of heat—is referred to
the author’s treatise on “Heat for Engineers,” issued by the publishers
of the present volume.
With regard to temperature scales, the author has in the main employed
Centigrade degrees, but has recognised that the Fahrenheit degree is
still largely used, and has therefore frequently expressed temperatures
in terms of both scales.
The number of those who find it an advantage in their calling to
measure and control high temperatures is constantly increasing; and
the manufacture of pyrometric appliances now gives employment to
considerable numbers. The author trusts that the present treatise will
prove of service to all thus concerned, and also to those who pursue
the fascinating study of high temperature measurement from the purely
scientific standpoint.
In conclusion, the author expresses his thanks to the various firms,
mentioned in the text, who have loaned blocks for the purpose of
illustration, and who have furnished him with much valuable information.
CHAS. R. DARLING.
WOOLWICH, 1920.
PYROMETRY
CHAPTER I
INTRODUCTION
The term “pyrometer”—formerly applied to instruments designed to
measure the expansion of solids—is now used to describe any device
for determining temperatures beyond the upper limit of a mercury
thermometer. This limit, in the common form, is the boiling point of
mercury: 357° C. or 672° F. By leaving the bore of the tube full of
nitrogen or carbon dioxide prior to sealing, the pressure exerted by
the enclosed gas when the mercury expands prevents boiling; and with a
strong bulb of hard glass the readings may be extended to 550° C. or
1020° F. Above this temperature the hardest glass is distorted by the
high internal pressure, but, by substituting silica for glass, readings
as high as 700° C. or 1290° F. may be secured. Whilst such thermometers
are useful in laboratory processes they are too fragile for workshop
use; and if made of the length necessary in many cases in which the
temperature of furnaces is sought, the cost would be as great as that
of more durable and convenient appliances. No other instrument, however,
is so simple to read as the thermometer; and for this reason it is
used whenever the conditions are favourable. The latest proposal in
this direction is due to Northrup, who has constructed a thermometer
containing tin enclosed in a graphite envelope, which is capable of
reading up to 1500° C. or higher. This instrument is described on page
216.
The origin and development of the science of pyrometry furnish
a notable example of the value of the application of scientific
principles to industry. Sir Isaac Newton was the first to attempt
to measure the temperature of a fire by observing the time taken to
cool by a bar of iron withdrawn from the fire; but, although Newton’s
results were published in 1701, it was not until 1782 that a practical
instrument for measuring high temperatures was designed. In that year
Josiah Wedgwood, the famous potter, introduced an instrument based on
the progressive contraction undergone by clay when baked at increasing
temperatures, which he used in controlling his furnaces, finding it
much more reliable than the eye of the most experienced workman. This
apparatus (described on page 211) remained without a serious rival for
forty years, and its use has not yet been entirely abandoned.
The next step in advance was the introduction of the expansion
pyrometer by John Daniell in 1822. The elongation of a platinum rod,
encased in plumbago, was made to operate a magnifying device, which
moved a pointer over a scale divided so as to read temperatures
directly. Although inaccurate as compared with modern instruments, this
pyrometer was the first to give a continuous reading, and required no
personal attention. The expansion pyrometer—with different expanding
substances—is still used to a limited extent.
The year 1822 was also marked by Seebeck’s discovery of
thermo-electricity. The generation of a current of electricity by
a heated junction of two metals, increasing with the temperature,
appeared to afford a simple and satisfactory basis for a pyrometer,
and Becquerel constructed an instrument on these lines in 1826.
Pouillet and others also endeavoured to measure temperatures by the
thermo-electric method, but partly owing to the use of unsuitable
junctions, and partly to the lack of reliable galvanometers, these
workers failed to obtain concordant results. The method was for all
practical purposes abandoned until 1886, when its revival in reliable
form led to the enormous extension of the use of pyrometers witnessed
during recent years.
In 1828 Prinsep initiated the use of gas pyrometers, and enclosed the
gas in a gold bulb. Later workers used porcelain bulbs, on account of
greater infusibility, but modern research has shown that porcelain is
quite unsuitable for accurate measurements, being porous to certain
gases at high temperatures, even when glazed. Gas pyrometers are
of little use industrially, but are now used as standards for the
calibration of other pyrometers, the bulb being made of an alloy of
platinum and rhodium.
Calorimetric pyrometers, based on Regnault’s “method of mixtures,”
were first made for industrial purposes by Byström, who patented an
instrument of this type in 1862. This method has been widely applied,
and a simplified form of “water” pyrometer, made by Siemens, is at
present in daily use for industrial purposes. It is not capable,
however, of giving results of the degree of accuracy demanded by many
modern processes.
The resistance pyrometer was first described by Sir W. Siemens in 1871,
and was made by him for everyday use in furnaces. Many difficulties
were encountered before this method was placed on a satisfactory
footing, but continuous investigation by the firm of Siemens & Co.,
and also the valuable researches of Callendar and Griffiths, have
resulted in the production of reliable resistance pyrometers, which are
extensively used at the present time.
In 1872 Sir William Barrett made a discovery which indirectly led to
the present development of the science of pyrometry. Barrett observed
that iron and steel, on cooling down from a white heat, suddenly became
hotter at a definite point, owing to an internal molecular change; and
gave the name of “recalescence” to the phenomenon. Workers in steel
subsequently discovered that this property was intimately connected
with the hardening of the metal; thus Hadfield noticed that a sample of
steel containing 1·16 per cent. of carbon, when quenched just below the
change-point was not hardened, but when treated similarly at 15° C.
higher it became totally hard. The demand for accurate pyrometers in
the steel industry followed immediately on these discoveries, for even
the best-trained workman could not detect with the eye a difference in
temperature so small, and yet productive of such profound modification
of the properties of the finished steel. In this instance, as in many
others, the instruments were forthcoming to meet the demand.
The researches of Le Chatelier, published in 1886, marked a
great advance in the progress of pyrometry. He discovered that a
thermo-electric pyrometer, satisfactory in all respects, could be
made by using a junction of pure platinum with a rhodioplatinum
alloy, containing 10 per cent. of rhodium; a d’Arsonval moving-coil
galvanometer being used as indicator. This type of galvanometer,
which permits of an evenly-divided scale, is now universally employed
for this purpose, and has made thermo-electric pyrometers not only
practicable, but more convenient for general purposes than any other
type. Continuous progress has since been made in connection with this
method, which is now more extensively used than any other.
Attempts to deduce temperature from the luminosity of the heated body
were first made by Ed. Becquerel in 1863, but the method was not
successfully developed until 1892, when Le Chatelier introduced his
optical pyrometer. This instrument, being entirely external to the
hot source, enabled readings to be taken at temperatures far beyond
the melting point of platinum, which would obviously be the extreme
limit of a pyrometer in which platinum was used. The quantitative
distribution of energy in the spectrum has since been worked out by
Wien and Planck, who have furnished formula based on thermodynamic
reasoning, by the use of which optical pyrometers may now be calibrated
in terms of the thermodynamic scale of temperature. Other optical
pyrometers, referred to in the text, have been devised by Wanner,
Holborn and Kurlbaum, Féry, and others; and the highest attainable
temperatures can now be measured satisfactorily by optical means.
The invention of the total-radiation pyrometer by Féry in 1902 added
another valuable instrument to those already available. Based on the
fourth-power radiation law, discovered by Stefan and confirmed by
the mathematical investigations of Boltzmann, this pyrometer is of
great service in industrial operations at very high temperatures,
being entirely external, and capable of giving permanent records.
Modifications have been introduced by Foster and others, and the method
is now widely applied.
Recorders, for obtaining permanent evidence of the temperature of a
furnace at any time, were first made for thermo-electric pyrometers by
Holden and Roberts-Austen, and for resistance pyrometers by Callendar.
Numerous forms are now in use, and the value of the records obtained
has been abundantly proved.
For scientific purposes, all pyrometers are made to indicate Centigrade
degrees, 100 of which represent the temperature interval between
the melting-point of ice and the boiling point of water at 760 mm.
pressure, the ice-point being marked 0° and the steam-point 100°.
In industrial life, however, the Fahrenheit scale is often used in
English-speaking countries, the ice-point in this case being numbered
32° and the steam-point 212°; the interval being 180°. A single degree
on the Centigrade scale is therefore 1·8 times as large as a Fahrenheit
degree, but in finding the numbers on each scale which designate a
given temperature, the difference in the zero position on the two
scales must be taken into account. When it is desired to translate
readings on one scale into the corresponding numbers on the other, the
following formula may be used:—
(C. reading) (F. reading - 32)
———————————— = ————————————————— .
5 9
Thus by substituting in the above expression, 660° C. will be found to
correspond to 1220° F. and 1530° F. to 832° C.
It is greatly to be regretted that all pyrometers are not made to
indicate in Centigrade degrees, as confusion often arises through the
use of the two scales. An agreement on this point between instrument
makers would overcome the difficulty at once, as the Centigrade scale
is now so widely used that few purchasers would insist on Fahrenheit
markings.
It may be pointed out here that no single pyrometer is suited to every
purpose, and the choice of an instrument must be decided by the nature
of the work in hand. A pyrometer requiring skilled attention should not
be entrusted to an untrained man; and it may be taken for granted that
to obtain the most useful results intelligent supervision is necessary.
In the ensuing pages the advantages and drawbacks of each type will be
considered; but in all cases it is desirable, before making any large
outlay on pyrometers, to obtain a competent and impartial opinion as
to the kind best suited to the processes to be controlled. Catalogue
descriptions are not always trustworthy, and instances are not wanting
in which a large sum has been expended on instruments which, owing to
wrong choice, have proved practically useless. An instrument suited to
laboratory measurements is often a failure in the workshop, and all
possibilities of this kind should be considered before deciding upon
the type of pyrometer to be used.
CHAPTER II
STANDARDS OF TEMPERATURE
=The Absolute or Thermodynamic Scale of Temperatures.=—All
practical instruments for measuring temperatures are based on some
progressive physical change on the part of a substance or substances.
In a mercury thermometer, the alteration in the volume of the liquid
is used as a measure of hotness; and similarly the change in volume
or pressure on the part of a gas, or the variation in resistance to
electricity shown by a metal, and many other physical changes, may be
employed for this purpose. In connection with the measurement of high
temperatures, many different physical principles are relied upon in the
various instruments in use, and it is of the greatest importance that
all should read alike under the same conditions. This result would not
be attained if each instrument were judged by its own performances. In
the case of a mercury thermometer, for example, we may indicate the
amount of expansion between the temperatures of ice and steam at 76
centimetres pressure, representing 100° Centigrade, by _a_; and then
assume that an expansion of 2_a_ will signify a temperature of 200°,
and so on in proportion. Similarly, we may find the increase in
resistance manifested by platinum between the same two fixed points,
and indicate it by _r_, and then assume that an increase of 2_r_ will
correspond to 200°. If now we compare the two instruments, we find
that they do not agree, for on placing both in a space in which the
platinum instrument registered 200°, the mercury thermometer would show
203°. A similar, or even greater, discrepancy would be observed if
other physical changes were relied upon to furnish temperature scales
on these lines, and it is therefore highly desirable that a standard
independent of any physical property of matter should be used. Such a
standard is to be found in the thermodynamic scale of temperatures,
originally suggested by Lord Kelvin. This scale is based upon the
conversion of heat into work in a heat engine, a process which is
independent of the nature of the medium used. A temperature scale
founded on this conversion is therefore not connected with any physical
property of matter, and furnishes a standard of reference to which all
practical appliances for measuring temperatures may be compared.[1]
When readings are expressed in terms of this scale, it is customary to
use the letter K in conjunction with the number: thus 850° K would mean
850 degrees on the thermodynamic scale.
When existing instruments are compared with this standard, it is found
that a scale based on the assumption that the volume of a gas free
to expand, or the pressure of a confined gas, increases directly as
the temperature is in close agreement with the thermodynamic scale.
It may be proved that if the gas employed were “perfect,” a scale in
exact conformity with the standard described would be secured; and
gases which approach nearest in properties to a perfect gas, such
as hydrogen, nitrogen, and air, may therefore be used to produce a
practical standard, the indications of which are nearly identical with
the thermodynamic scale. If any other physical change be chosen, such
as the expansion of a solid, or the increase in resistance of a metal,
and a temperature scale be based on the supposition that the change
in question varies directly as the temperature, the results obtained
would differ considerably from the absolute standard. For this reason
the practical standard of temperature now universally adopted is an
instrument based on the properties of a suitable gas.
=The Constant Volume Gas Thermometer.=—In applying the properties
of a gas to practical temperature measurement, we may devise some
means of determining the increase in volume when the gas is allowed to
expand, or the increase in pressure of a confined gas may be observed.
The latter procedure is more convenient in practice, and the instrument
used for this purpose is known as the constant volume gas thermometer,
one form of which is shown in fig. 1. The gas is enclosed in a bulb B,
connected to a tube bent into a parallel branch, into the bend of which
is sealed a tap C, furnished with a drying cup. The extremity of the
parallel branch is connected to a piece of flexible tubing T, which
communicates with a mercury cistern which may be moved over a scale,
the rod G serving as a guide. In using this instrument the bulb B
is immersed in ice, and the tap C opened. When the temperature has
fallen to 0° C., the mercury is brought to the mark A by adjusting
the cistern, and the tap C then closed. The bulb B is now placed in
the space or medium of which the temperature is to be determined, and
expansion prevented by raising the cistern so as to keep the mercury
at A. When steady, the height of the mercury in the cistern above the
level of A is read off, and furnishes a clue to the temperature of B.
If the coefficient of pressure of the gas used (in this case, air) be
known, the temperature may be calculated from the equation
P_{1} = P_{0}(1 + _bt_),
where P_{1} is the pressure at _t_°; P_{0} the pressure at 0°; and _b_
the coefficient of pressure; that is, the increase in unit pressure
at 0° for a rise in temperature of 1°. Thus if P_{0} = 76 cms.; _b_ =
0·00367; height of mercury in cistern above A = 55·8 cms.; then
P_{1} = (76 + 55·8) = 131·8 cms.,
and by inserting these values in the above equation _t_ is found to be
200°. In the instrument described, P_{0} is equal to the height of the
barometer, since the tap C is open whilst the bulb is immersed in ice.
The coefficient of pressure may be determined by placing the bulb in
steam at a known temperature, and noting the increased pressure. In the
equation given, P_{1}, P_{0}, and _t_ are then known, and the value of
_b_ may be calculated.
[Illustration: FIG. 1.—CONSTANT VOLUME AIR THERMOMETER.]
In using this instrument for exact determinations of temperature,
allowance must be made for the expansion of the bulb, which causes a
lower pressure to be registered than would be noted if the bulb were
non-expansive. Again, the gas in the connecting tube is not at the same
temperature as that in the bulb; an error which may be practically
eliminated by making the bulb large and the bore of the tube small.
The temperature of the mercury column must also be allowed for, as the
density varies with the temperature. When the various corrections have
been made, readings of great accuracy may be secured.
When applied to the measurement of high temperatures, the bulb must
be made of a more infusible material than glass. Gold, porcelain,
platinum, and quartz have been used by different investigators, but
the most reliable material for temperatures exceeding 900° C. has
been found to be an alloy of platinum with 20 per cent. of rhodium.
The most suitable gas to use inside the bulb is nitrogen, which is
chemically inert towards the materials of the bulb, and is not absorbed
by the metals mechanically. When measuring high temperatures with this
instrument, a considerable pressure, amounting to 1 atmosphere for
every increase of 273 degrees above the ice point, is requisite to
prevent expansion of the nitrogen; and this pressure tends to distort
the bulb and so to falsify the indications. This trouble has been
overcome by Day, who surrounded the bulb by a second larger bulb, and
forced air or nitrogen into the intervening space until the pressure
on the exterior of the thermometer bulb was equal to that prevailing
in the interior. Even then it was not found possible to secure higher
readings than 1550° C., as the bulb commenced to alter in shape owing
to the softening of the material. This temperature represents the
highest yet measured on the gas scale; but by using a more refractory
material, such as fused zirconia, it may be found possible to extend
this range to 2000° C. or more. Experiments in this direction are
very desirable, in order that high-reading pyrometers may be checked
directly against the gas scale.
=Fixed Points for Calibration of Pyrometers.=—It is evident
that the gas thermometer is totally unsuited for use in workshops
or laboratories when a rapid determination of a high temperature is
required. Its function is to establish fixed points or temperature
standards, by means of which other instruments, more convenient to
use, may be graduated so as to agree with each other and with the gas
scale itself. The temperature scales of all modern pyrometers are thus
derived, directly or indirectly, from the gas thermometer. In the
table on next page, a number of fixed points, determined by various
observers, is given; the error, even at the highest temperatures,
probably not exceeding ±2° C.
In preparing the temperature scale of a pyrometer for practical
use, the instrument is subjected successively to a number of the
temperatures indicated in the table, and in this manner several fixed
points are established on its scale. The space between these points is
then suitably subdivided to represent intermediate temperatures.
TABLE OF FIXED POINTS.
──────────────────┬────────────────────┬───────┬──────
Substance. │Physical Condition. │ Deg. │ Deg.
│ │ Cent. │ Fahr.
──────────────────┼────────────────────┼───────┼───────
Water (ice) │ At Melting Point │ 0 │ 32
Water │ ” Boiling ” │ 100 │ 212
Aniline │ ” ” ” │ 184 │ 363
Naphthalene │ ” ” ” │ 218 │ 424
Tin │ ” Melting ” │ 232 │ 449
Lead │ ” ” ” │ 327 │ 620
Zinc │ ” ” ” │ 419 │ 786
Sulphur │ ” Boiling ” │ 445 │ 833
Antimony │ ” Melting ” │ 631 │ 1167
Aluminium │ ” ” ” │ 657 │ 1214
Common Salt │ ” ” ” │ 800 │ 1472
Silver (in air) │ ” ” ” │ 955 │ 1751
Silver (free from │ │ │
oxygen) │ ” ” ” │ 962 │ 1763
Gold │ ” ” ” │ 1064 │ 1947
Copper (in air) │ ” ” ” │ 1064 │ 1947
Copper (Graphite │ │ │
covered) │ ” ” ” │ 1084 │ 1983
Iron (pure) │ ” ” ” │ 1520 │ 2768
Palladium │ ” ” ” │ 1549 │ 2820
Platinum │ ” ” ” │ 1755 │ 3190
──────────────────┴────────────────────┴───────┴───────
It is necessary to point out that the figures given in the table refer
only to pure substances, and that relatively small quantities of
impurities may give rise to serious errors. The methods by which the
physical condition to which the temperatures refer may be realised in
practice will be described in the succeeding chapter.
=National Physical Laboratory Scale.=—Exact agreement with regard
to fixed points has not yet been arrived at in different countries, and
an effort to co-ordinate the work of the National Physical Laboratory,
the United States Bureau of Standards, and the Reichsanstalt, with a
view to the formation of an international scale, was interrupted by the
war. In 1916 the National Physical Laboratory adopted a set of fixed
points on the Centigrade thermodynamic scale, in conformity with which
all British pyrometers have since been standardised. It will be seen
that the figures differ very slightly from those given in the previous
table, which represent the average results of separate determinations
in different countries.
NATIONAL PHYSICAL LABORATORY SCALE (1916).
────────────────────┬──────────────────────────┬────────┬──────
Substance. │ Physical Condition. │ Deg. │ Deg.
│ │ Cent. │ Fahr.
────────────────────┼──────────────────────────┼────────┼──────
Water (ice) │ At Melting Point │ 0 │ 32
Water │ ” Boiling ” (760 mm.)│ 100 │ 212
Naphthalene │ ” ” ” ” │ 217·9 │ 424
Benzophenone │ ” ” ” ” │ 305·9 │ 582
Zinc │ ” Melting ” │ 419·4 │ 787
Antimony │ ” ” ” │ 630 │ 1166
Common Salt │ ” ” ” │ 801 │ 1474
Silver (in reducing │ │ │
atmosphere) │ ” ” ” │ 961 │ 1761
Gold │ ” ” ” │ 1063 │ 1945
Copper (in reducing │ │ │
atmosphere) │ ” ” ” │ 1083 │ 1982
────────────────────┴──────────────────────────┴────────┴──────
For higher temperatures the melting points of nickel (1452° C.) and
palladium (1549° C.) are employed, but the accuracy in these cases is
not so certain as with the substances named in the table. A useful
point, intermediate between copper and nickel, has been established
by E. Griffiths, and is obtained by heating nickel with an excess of
graphite, when a well-defined eutectic is formed which freezes at 1330°
C., or 2426° F.
=Temperatures above the Present Limit of the Gas Thermometer.=—As
it is not yet possible to compare an instrument directly with the
gas thermometer above 1550° C., all higher temperatures must be
arrived at by a process of extrapolation. By careful observation of a
physical change at temperatures up to the limit of 1550° C., the law
governing such change may be discovered; and assuming the law to hold
indefinitely, higher temperatures may be deduced by calculation. An
amount of uncertainty always attaches to this procedure, and in the
past some ludicrous figures have been given as the result of indefinite
extrapolation. Wedgwood, for example, by assuming the uniform
contraction of clay, gave 12001° C., or 21637° F., as the melting point
of wrought iron, whereas the correct figure is 1520° C., according
to the gas scale. Even in recent times, the extrapolation of the law
connecting the temperature of a thermal junction with the electromotive
force developed, obtained by comparison with the gas scale up to 1100°
C., led Harker to the conclusion that the melting point of platinum
was 1710° C., a figure 45 degrees lower than that now accepted. The
laws governing the radiation of energy at different temperatures,
however, appear to be capable of mathematical proof from thermodynamic
principles, and temperatures derived from these laws are in reality
expressed on the absolute or thermodynamic scale. Extrapolation of
these laws, when used to deduce temperatures by means of radiation
pyrometers, appears to be justified; but it is still desirable to
extend the gas scale as far as possible to check such instruments.
Assuming the radiation laws to hold, it is possible to determine the
highest temperatures procurable, such as that of the electric arc, with
a reasonable degree of certainty.
[1] For a fuller account of the thermodynamic scale, see the author’s
treatise _Heat for Engineers_, pp. 391-2.
CHAPTER III
THERMO-ELECTRIC PYROMETERS
=General Principles.=—Seebeck, in 1822, made the discovery that
when a junction of two dissimilar metals is heated an electromotive
force is set up at the junction, which gives rise to a current
of electricity when the heated junction forms part of a closed
circuit. Becquerel, in 1826, attempted to apply this discovery to
the measurement of high temperatures, it having been observed that
in general the E.M.F. increased as the temperature of the junction
was raised. No concordant results were obtained, and the same fate
befell the investigations of others who subsequently attempted to
produce pyrometers based on the Seebeck effect. These failures were
due to several causes, but chiefly to the non-existence of reliable
galvanometers, such as we now possess. It was not until 1886 that the
problem was satisfactorily solved by Le Chatelier of Paris.
Although any heated junction of metals will give rise to an
electromotive force, it does not follow that any pair, taken at random,
will be suited to the purposes of a pyrometer. A junction of iron and
copper, for example, gives rise to an E.M.F. which increases with the
temperature up to a certain point, beyond which the E.M.F. falls off
although the temperature rises, and finally reverses in direction—a
phenomenon to which the name of “thermo-electric inversion” has been
applied. Evidently, it would be impossible to measure temperatures
in this case from observations of the electromotive force produced,
and any couple chosen must be free from this deterrent property.
Moreover, the metals used must not undergo deterioration, or alteration
in thermo-electric properties, when subjected for a prolonged
period to the temperature it is desired to measure. These and other
considerations greatly restrict the choice of a suitable pair of
metals, which, to give satisfaction, should conform to the following
conditions:—
1. The E.M.F. developed by the junction should increase uniformly as
the temperature rises.
2. The melting point of either component should be well above the
highest temperature to be measured. An exception to this rule
occurs when the E.M.F. of fused materials is employed.
3. The thermo-electric value of the couple should not be altered by
prolonged heating.
4. The metals should be capable of being drawn into homogeneous
wires, so that a junction, wherever formed, may always give rise
to the same E.M.F. under given conditions.
It is a further advantage if the metals which fulfil the above
conditions are cheap and durable.
The exacting character of these requirements delayed the production
of a reliable thermo-electric pyrometer until 1886, when Le Chatelier
discovered that a junction formed of platinum as one metal, and an
alloy of 90 per cent. of platinum and 10 per cent. of rhodium as the
other, gave concordant results. In measuring the E.M.F. produced, Le
Chatelier took advantage of the moving-coil galvanometer introduced by
d’Arsonval, which possessed the advantages of an evenly-divided scale
and a dead-beat action. This happy combination of a suitable junction
with a simple and satisfactory indicator immediately established the
reliability of the thermo-electric method of measuring temperatures.
As platinum melts at 1755° C., and the rhodium alloy at a still higher
temperature, a means was thus provided of controlling most of the
industrial operations carried out in furnaces.
So far, the effect of heating the junction has been considered without
regard to the temperature of the remainder of the circuit, and it is
necessary, before describing the construction of practical instruments,
to consider the laws governing the thermo-electric circuit, the
simplest form of which is represented in fig. 2. One of the wires is
connected at both ends to separate pieces of the other wire, the free
ends of which are taken to the galvanometer Two junctions, A and B, are
thus formed, which evidently act in opposition; for if on heating A the
direction of current be from A to B, then on heating B the direction
will be from B to A. Hence if A and B were equally heated no current
would flow in the circuit, the arrangement being equivalent to two
cells of equal E.M.F. in opposition. Thermal junctions are formed at
each of the galvanometer terminals, but the currents to which they give
rise, when the temperature changes, are opposed and cancel each other.
The law which holds for this circuit may be expressed thus:—
“If in a thermo-electric circuit there be two junctions, A and B,
the electromotive force developed is proportional to the
_difference_ in temperature between A and B.”
[Illustration: FIG. 2.—TWO-JUNCTION THERMO-ELECTRIC CIRCUIT.]
It is customary to refer to the two junctions as the “hot” and “cold”
junctions; but it is important to remember that fluctuations in the
temperature of either will alter the reading on the galvanometer or
indicator.
A second law, which applies to all thermo-electric circuits, is that
“the E.M.F. developed is independent of the thickness of the wire.”
This does not mean that the deflection of the galvanometer is the
same whether thin or thick wires are used to form the junction. The
deflection depends upon the current flowing through the circuit, and
this, according to Ohm’s law, varies inversely as the total resistance
of the circuit. Consequently, the use of thin wires of a given kind
will tend to give a less deflection than in the case of thick wires, as
the resistance of the former will be greater, and unless the resistance
of the galvanometer be great compared with that of the junction, the
difference in deflection will be conspicuous. The E.M.F., however, is
the same under given conditions, whatever thickness of wire be used.
Reference to fig. 2 will show that in order to realise this circuit in
practice, one of the wires forming the couple must be used in the form
of leads to the galvanometer. This can readily be done if the material
of the wire is cheap; but if platinum or other expensive metal be
used, and the galvanometer be some yards distant, the question of cost
necessitates a compromise, and the circuit is then arranged as in fig.
3. The wires forming the hot junction are brought to brass terminals T
T, from which copper wires lead to the galvanometer G. This arrangement
results in three effective junctions, viz. the hot junction A to B; the
junction A to brass, and the junction B to brass. It will be seen that
the two junctions of copper to brass are in opposition, and cancel
each other for equal heating; and the same applies to the galvanometer
connections. A circuit thus composed of three separate junctions does
not permit of a simple expression for the net E.M.F. under varying
temperature conditions, and to avoid errors in readings care must be
taken to prevent any notable change of temperature at the terminals T T
in a practical instrument arranged as in the diagram.
[Illustration: FIG. 3.—THREE-JUNCTION THERMO-ELECTRIC CIRCUIT.]
A point of practical utility in thermo-electric work is the fact that
if a wire be interrupted by a length of other metal, as indicated at C
in fig. 3, no current will be set up in a circuit if both joints are
equally heated, as the electromotive forces generated at each junction
are in opposition. It is thus possible to interrupt a circuit by a
plug-key or switch, without introducing an error; always provided that
an even temperature prevails over the region containing the joints.
Another useful fact is that if two wires be brought into contact, they
may be fastened over the joint by soldering or using a third metal,
without alteration of thermo-electric value, except in rare cases. Thus
a copper-constantan or iron-constantan junction may suitably be united
by silver solder, using borax as a flux, thus avoiding the uncertainty
of contact which must always occur when the wires are merely twisted
together. Welding, however, is preferable to soldering.
=Metals used for Thermal Junctions.=—Until recent years
it was customary to employ a platinum-rhodioplatinum or
platinum-iridioplatinum junction for all temperatures beyond the
scope of the mercury thermometer. The almost prohibitive price of
these metals has caused investigations to be made with a view to
discovering cheaper substitutes, with successful results up to 1000°
C. or 1800° F., thus comprehending the range of temperatures employed
in many industrial processes. Above this temperature the platinum
series of metals are still used for accurate working, but it will be
of great advantage if the range measurable by cheap or “base” metals
can be further extended. Promise in this direction is afforded by
the properties of fused metals when used in thermal junctions. An
investigation by the author has shown that in general the E.M.F.
developed by a junction does not undergo any sudden change when one
or both metals melt, but continues as if fusion had not occurred. By
making arrangements to maintain the continuity of the circuit after
fusion, it may be possible to read temperatures approximating to the
boiling points of metals such as copper and tin, both of which are over
2000° C. The base metals are not so durable as platinum and kindred
metals, but as the cost of replacement is negligible, this drawback is
of little importance. Moreover, base-metal junctions usually develop a
much higher E.M.F. than the platinum metals, which enables stronger and
cheaper galvanometers to be used as indicators.
THERMAL JUNCTIONS USED IN PYROMETERS.
─────────────────────────────────────────────────┬────────────────
│ Upper limit to
│ which Junction
│ may be used.
Couple. ├────────┬───────
│ Deg. │ Deg.
│ Cent. │ Fahr.
─────────────────────────────────────────────────┼────────┼───────
Platinum and rhodioplatinum (10 per cent. Rh) │ 1400 │ 2550
2 Rhodioplatinum alloys of different composition │ 1600 │ 2900
Platinum and iridioplatinum (10 per cent. Ir) │ 1100 │ 2000
Nickel and constantan │ 900 │ 1650
Nickel and copper │ 800 │ 1475
Nickel and carbon │ 1000 │ 1850
Nickel and iron │ 1000 │ 1850
Iron and constantan │ 900 │ 1650
Copper and constantan │ 800 │ 1475
Silver and constantan │ 800 │ 1475
2 Nickel chrome alloys of different composition │ │
(Hoskin’s alloys) │ 1100 │ 2010
Nickel-chrome alloy and nickel-aluminium alloy │ 1100 │ 2010
2 Iron-nickel alloys of different composition │ 1000 │ 1850
─────────────────────────────────────────────────┴────────┴───────
The electromotive force developed by a junction of any given pair of
metals when heated to a given temperature varies according to the
origin of the metals. It is not unusual, for example, for two samples
of 10 per cent. rhodioplatinum, obtained from different sources, to
show a difference in this respect of 40 per cent. when coupled with
the same piece of platinum. Equal or greater divergences may be noted
with other metals; and hence the replacement of a junction can only be
effected, with accuracy, by wires from the same lengths of which the
junction formed a part. As showing how platinum itself is not uniform,
it may be mentioned that almost any two pieces of platinum wire, if
not from the same length, will cause a deflection on a sensitive
galvanometer when made into a junction and heated. It is therefore
customary for makers to obtain considerable quantities of wire of a
given kind, homogeneous as far as possible, in order that a number of
identical instruments may be made, and the junctions replaced, when
necessary, without alteration of the scale of the indicator.
The alloy known as “constantan,” which figures largely in the foregoing
table, is composed of nickel and copper, and is practically identical
with the alloy sold as “Eureka” or “Advance.” It has a high specific
resistance, and a very small temperature coefficient, and is much used
for winding resistances. Couples formed of constantan and other metals
furnish on heating an E.M.F. several times greater than that yielded
by couples of the platinum series, and show an equally steady rise
of E.M.F. with temperature. This alloy has proved of great service in
connection with the thermo-electric method of measuring temperatures.
Couples formed of nickel-chrome alloys, known as “Hoskin’s alloys,”
have been introduced into Britain by the Foster Instrument Company,
which may be used continuously to 1100° C., and for occasional readings
up to 1300° C. Another couple, much used in America, consists of an
alloy of 90 per cent. nickel and 10 per cent. chromium, and an alloy
of 98 per cent. nickel and 2 per cent. aluminium, which may be used up
to 1100° C. Other couples, formed of alloys of nickel, chromium, iron,
aluminium, etc., have been introduced by different makers, but have not
proved so satisfactory as those mentioned above.
=Changes in Thermal Junctions when constantly used.=—No metal
appears to be able to withstand a high temperature continuously without
undergoing some physical alteration; and for this reason the E.M.F.
developed by a given junction is liable to change after a period of
constant use. At temperatures above 1100° C., platinum, for example,
undergoes a notable change in a comparatively short period, but below
1000° C., the change is very slight, and if this range be not exceeded,
a platinum-rhodioplatinum or iridioplatinum junction may be used for
years without serious error arising from this cause. This liability
to change is one of the factors which restricts the range of thermal
junctions, which should never be used continuously beyond the
temperature at which the alteration commences to become large. A second
cause of discrepancy is the possible alteration in the composition
of an alloy, due to one of the constituents leaving in the form of
vapour, as is noted with iridioplatinum alloys, from which the iridium
volatilises in tangible quantities above 1100° C., causing a fall of
10 per cent. or more in the thermo-electric value of the junction of
these alloys with platinum. Constantan appears to be very stable in
its thermo-electric properties, and the various junctions in which
it plays a part show a high degree of stability if not overheated.
Rhodioplatinum alloys are very stable, and for temperatures exceeding
1100° C. a junction of two of these alloys, of different composition,
is more durable than one in which pure platinum is used. An extended
series of tests on base-metal junctions made in America by Kowalke
showed that continuous heating of couples as received from the makers
altered the E.M.F. considerably, the change in some cases representing
over 100° C. on the indicator. A stable condition, due to the relief of
strains or other change, was finally reached, and the conclusion drawn
that the materials should be thoroughly annealed before calibration. It
is desirable in all cases periodically to test the junctions at some
standard temperature, and if any conspicuous error be noted, to replace
the old junction by a new one.
In addition to the errors due to slow physical changes, a junction
may be altered considerably, if imperfectly protected, owing to the
chemical action of furnace gases, or of solids with which the junction
may come into contact. The vapours of metals such as lead or antimony
are very injurious; and platinum in particular is seriously affected
by vapours containing phosphorus, if in a reducing atmosphere. So
searching is the corrosive action of furnace gases that adequate
protection of the junction is essential if errors and damage are to be
avoided. When a wire has once been corroded, a junction made with it
will not develop the same E.M.F. as before.
=Electromotive Force developed by Typical Junctions.=—The
following table exhibits the E.M.F. generated by several junctions
for a range of 100° C., taken at the middle part of the working range
in each case. These figures are subject to considerable variation,
according to the origin of the metals.
──────────────────────────────────────────┬─────────────────────────
│ E.M.F. in millivolts for
Couple. │ a rise of 100° at middle
│ of working range.
──────────────────────────────────────────┼─────────────────────────
Platinum-rhodioplatinum (10 per cent. Rh) │ 1·1
Platinum-iridioplatinum (10 per cent. Ir) │ 1·2
Nickel-constantan │ 2·3
Copper-constantan │ 5·8
Nickel-copper │ 6·1
Iron-constantan │ 6·7
Hoskin’s alloys │ 7·4
──────────────────────────────────────────┴─────────────────────────
It will be noted that the base-metal junctions give much higher values
than the platinum series, and hence can be used with a less sensitive,
and therefore cheaper, indicator. Base-metal junctions are also, in
consequence of the greater E.M.F. furnished, capable of yielding more
sensitive readings over a selected range of temperature.
[Illustration: FIG. 4.—PRACTICAL FORM OF THERMO-ELECTRIC PYROMETER.]
=Practical Forms of Thermocouples.=—When expensive junctions are
employed, wires of the minimum thickness consistent with strength and
convenience of construction are used, a diameter of No. 25 standard
wire gauge being suitable. A common arrangement is shown in fig. 4, in
which J is the hot junction, the wires from which are passed through
thin fireclay tubes which serve as insulators (or through twin-bore
fireclay) to the reels R R, in the head of the pyrometer, upon which
a quantity of spare wire is wound to enable new junctions to be made
when required. Two brass strips, S, are screwed down on to the wires
at one end, and are furnished with screw terminals at the other
end, from which wires are taken to the galvanometer or indicator. A
protecting-tube, T, surrounds the wires and hot junction. The head,
H, may be constructed of wood, fibre, or porcelain, and should be an
insulator for electricity and heat. There are various modifications
in use, but the general method described is adopted by most makers.
In order to guard against errors arising from alterations in the
temperature of the cold junctions in the end of the pyrometer, some
firms construct the head so as to leave a hollow space, through which
cold water is constantly circulated (fig. 5), the arrangement being
known as a “water-cooled head.” In some forms the supply of spare wires
is made to take the form of two spiral springs in a hollow head, the
upper ends of the springs being taken to terminals.
[Illustration: FIG. 5.—PYROMETER WITH WATER-COOLED HEAD.]
The choice of a protecting-tube is a matter of considerable importance.
Obviously, such a tube should not soften at the highest temperature
attained, and when expensive metals are used to form the junction the
sheath should not be permeable to gases or vapours. It should also, if
possible, be a good conductor of heat, so that the junction may respond
quickly to a change of temperature in its surroundings, and should be
mechanically strong. It is difficult to secure all these properties
in any single material, and the choice of a sheath is decided by
the conditions under which the couple is to be used. The substances
employed, and their properties and special uses, may be enumerated as
follows:—
1. _Iron or Mild Steel._—For temperatures not exceeding 1100° C.
iron or mild steel covers are cheap and efficient from the standpoint
of conductivity, although liable to deteriorate owing to oxidation.
The tendency to oxidise is greatly diminished by “calorising” the
exterior by Ruder’s process, in which the iron is heated in a mixture
of metallic aluminium and oxide of aluminium, a surface alloy being
formed which resists oxidation. A result nearly as good may be obtained
by smearing the surface with fine aluminium powder, and bringing to a
white heat. This treatment greatly prolongs the life of an iron sheath.
Some makers employ an inner steel tube round the wires, and an outer
tube which comes into contact with the furnace gases, corrosion of the
latter being detected before the inner tube has given way and exposed
the junction. Some makers do not consider it safe to expose heated
platinum to an iron surface, with only air intervening, and hence use
an inner cover of silica or porcelain, which the outer iron or steel
tube protects from mechanical damage. For ordinary work seamless steam
or hydraulic steel tubing, with a welded end, is satisfactory; but for
dipping into molten lead or other metals the tube should be bored
from the solid. The great advantage of an iron or steel sheath is its
mechanical strength, which protects the couple from damage in case of
rough usage.
2. _Nichrom._—Certain alloys of nickel and chromium, and especially
that known as Nichrom II, may be kept at 1100° C. without oxidising
to any appreciable extent; and hence sheaths of this material may be
used up to the temperature named. In addition to being more durable
than iron, nichrom possesses the same advantages of strength and good
conductivity; on the other hand, it is more costly.
3. _Molybdenum._—This metal, which possesses a melting point of
about 2500° C., may be dipped in molten brass, bronze, copper, etc.,
without being attacked, and has been used to form the tip of a
protecting-tube designed to measure the temperature of molten alloys. A
junction covered only by a thin tube of molybdenum quickly attains the
temperature of its surroundings.
4. _Graphite and Graphite Compositions._—Carbon has the highest
melting point of all known substances, and in the form of artificial or
Acheson graphite may be easily machined to any desired shape. Graphite
sheaths are sometimes used for immersion in molten metals, but at 1000°
C. and higher Acheson graphite oxidises easily and becomes friable. It
is a good conductor of heat, but is easily broken. Compositions of
natural graphite and refractory earths, such as Morgan’s “Salamander,”
are inferior to pure graphite in conductivity, but are stronger and not
readily oxidised, and may be used to form sheaths for temperatures up
to 1400° C. or possibly higher, when penetration of furnace gases to
the junction is not of moment.
5. _Porcelain._—This material, in its best forms, may be used up to
1400° C., but must be efficiently glazed to prevent the ingress of
furnace gases to the junction. It is easily broken by a blow, and when
circumstances permit should be protected by an iron covering-sheath.
The variety known as “Marquardt” has been found very satisfactory for
high-reading thermal couples. Porcelain is not a good conductor of
heat, and a junction encased in it does not respond quickly to external
changes in temperature.
6. _Vitrified Silica._—This substance, which may be worked in the
oxy-hydrogen blowpipe, is largely used as a protecting-tube. It is not
advisable, however, to use it for continuous work above 1100° C., as
beyond this temperature devitrification occurs, and the tube becomes
porous. It is a fairly good conductor of heat, and withstands rapid
changes in temperature without cracking. It is very brittle, and for
this reason is generally encased in iron.
7. _Alundum._—This material is made from fused bauxite, and
has a melting point of 2050° C. A special form of alundum, used
for protecting-tubes, is non-porous up to 1300° C., and forms a
satisfactory covering. Alundum is a moderately good conductor of heat,
but is easily broken.
8. _Carborundum._—This is an electric furnace product, which may
be heated above 2000° C. without damage. For making into pyrometer
tubes, it is bonded with a suitable material, and baked after shaping.
Carborundum, and the amorphous variety known as “silfrax,” have proved
useful for protecting junctions at temperatures as high as 1600° C.
The thermal conductivity is relatively good, but the tubes are easily
broken.
9. _Magnesia._—Tubes of this material, which melts at a temperature
considerably above 2000° C., have been used for special work. Magnesia
is a poor conductor of heat, and has little mechanical strength.
10. _Zirconia._—This is a very refractory material, its melting point
exceeding 2500° C. It may be made into a vitreous variety, which is
non-porous and proof against sudden temperature changes. At present,
only a moulded form of pyrometer tube, made from zirconia powder, is
available, the material worked in this manner being termed “zirkite.”
Although zirconia is a bad conductor of heat, its other qualities
are such that it forms an excellent material for work at the highest
temperatures possible for thermal junctions; and when the vitreous
variety is available, may come into extended use.
[Illustration: FIG. 6. PYROMETER WITH SPECIAL COLD JUNCTION IN HEAD.]
It will be seen from the foregoing that the ideal protecting-tube has
yet to be found, and the user must choose the one which comes nearest
to his requirements. Special consideration must be given in cases when
chemical fumes are present, and a sheath selected which is not attacked
or penetrated by them.
Returning to the junction, it is advisable always to weld the wires,
and not to rely upon the contact resulting from twisting them together.
Platinum and the platinum alloys may be welded readily by placing the
junction in a coal-gas blowpipe fed with oxygen instead of air. For
work at lower temperatures the platinum metals may be soldered by means
of a small quantity of gold, in the flame of a Bunsen burner.
When cheap metals are used for the junction the construction may
be considerably modified, and often with advantage. In fig. 6, for
example, which represents a thermocouple made by A. Gallenkamp & Co.,
the metals used are copper and constantan, and the hot junction,
fastened by silver solder, is supplemented by a cold junction of the
same metals located in the head. The copper wire from the hot junction
passes directly to a copper terminal, from whence a copper wire lead
is carried to the galvanometer; and the same procedure is carried out
with the copper wire from the cold junction, thus realising the circuit
shown in fig. 2. The cold junction is kept in oil, the temperature
of which is registered by a short thermometer, thus enabling (as
will be explained later) the correct temperature of the hot junction
to be deduced under any circumstances. In this instrument twin-bore
fireclay is used to insulate the wires, and the protecting-tube is
of iron—which suffices for the upper limit (800° C.) to which the
junction may be used. Iron and constantan could be used in this manner
by employing iron leads to the galvanometer.
Another type of instrument, rendered practicable by the use of cheap
metals, and which may be termed the “heavy type,” is constructed of
thick pieces of the metals welded together instead of wires, thus
ensuring greater strength and longer life. Messrs Crompton & Co. were
the first to introduce thermocouples of this type, consisting of a
heavy steel tube, to one end of which a nickel rod is welded, the other
end being free, and the length of the rod suitably insulated from the
steel tube; leads for the rod and tube being taken to the galvanometer.
Fig. 7 shows a couple of this kind, made by Paul, consisting of an iron
tube down the middle of which a constantan rod is passed, insulated
from the tube by magnesia. At the tapered end the two metals are welded
together, and at the free end a special cap, fitted over the tube and
rod, the contact parts being insulated from one another, serves to
enable leads to be taken to the galvanometer. Similar thermocouples are
made by the Foster Instrument Company (fig. 8), and are simple, cheap,
and reliable up to 900° C. with an iron-constantan couple, and to 1100°
C. with nichrom couples. When worn out they may be replaced, at a
trifling cost, by others made from the same batch of metal.
[Illustration: FIG. 7.—HEAVY TYPE, CHEAP-METAL PYROMETER.]
The drawback to the use of carbon as one of the materials for a
junction is the difficulty experienced in securing a good contact with
the metal with which it is coupled. In nickel-carbon junctions the
contact is sometimes ensured by the aid of a spring, which presses
the two substances together. Such an arrangement is evidently not so
reliable as one in which the materials are welded, and a defective
contact, arising from any cause, would lead to serious error. A
preferable plan is to screw both the nickel and carbon rods into a
cross-piece of either element.
[Illustration: FIG. 8.—FOSTER’S CHEAP-METAL PYROMETER.]
When applying a thermal junction to the measurement of surface
temperatures, such as steam-pipes or the exterior of furnaces, the
wires may be passed through a thin disc of metal, about ¼ in. in
diameter, and soldered at the back. Suitable materials are copper and
constantan, soldered to a thin copper disc with silver solder, and
brought to a cold junction in the head of the instrument as shown in
fig. 6. The terminal piece of the insulation may be made of hard wood,
with the holes countersunk so as to cover the solder and enable the
wood to touch the disc, which, when pressed on the hot surface, will
then rapidly acquire the temperature. The author has found, by trials
under varying circumstances, that this method of measuring surface
temperatures gives reliable and concordant results. For very high
surface temperatures a platinum disc, with one of the usual platinum
metal couples soldered to the disc with pure silver, and a piece of
twin-bore fireclay brought to the back of the disc, will be found to
suffice for most cases arising in practice. A small blowpipe flame is
best for soldering the wires to the disc, borax being used as flux in
the first case; but no flux is necessary in soldering the platinum
metals with pure silver.
In deciding upon the length of a thermocouple it must be remembered
that the temperature recorded is that prevailing in the region of
the hot junction. When the temperature of a furnace is uniform it is
sufficient to allow the end of the thermocouple to protrude about 12
inches into the interior, but when following the change of temperature
undergone by objects in a furnace the end must be located near the
objects. If the distance from the exterior of the furnace to the
objects exceed 2 feet, the thermocouple should be inserted through the
roof so as to hang vertically, as if placed through the side it would
droop by its own weight at high temperatures. The distance between
the exterior of the furnace and the cold junctions should be at least
15 inches in all cases in which the heating of the cold junction
is not automatically compensated. After inserting the couple the
opening through the furnace wall should be closed by means of suitable
luting-clay.
In certain instances, such as flues, it is necessary to use a long
instrument in a horizontal position. A rail may then be placed across
the flue, at a suitable place, to serve as a support and so to prevent
drooping.
=Liquid Element Thermocouples.=—An investigation by the author
and A. W. Grace has shown that the continuity of the E.M.F. produced
by a rising temperature is not interrupted by fusion, except in the
cases of bismuth and antimony, which both show an abrupt change in
thermo-electric properties at the melting point. It would therefore
appear feasible to measure temperatures by constructing a thermocouple
so as to retain the circuit after fusion, the advantage gained being
that the range is restricted by the boiling point of the metals instead
of the melting point and higher readings are rendered possible. The
boiling points of some of the common metals are appended:—
───────────┬───────────────────
Metal. │ Boiling Point.
───────────┼─────────┬─────────
│ Deg. C. │ Deg. F.
├─────────┼─────────
Aluminium │ 1800 │ 3270
Silver │ 1955 │ 3550
Tin │ 2270 │ 4120
Copper │ 2310 │ 4190
Nickel │ 2330 │ 4225
Iron │ 2450 │ 4440
───────────┴─────────┴─────────
From inspection of these figures, it will be seen that if a suitable
couple could be obtained, common metals might be used to measure
temperatures equalling or even exceeding the limit of the range covered
by wire junctions of metals of the platinum series. Instead of using
two metals, graphite might form one member of the couple, provided that
no objection to its use existed on other grounds.
[Illustration: FIG. 9.—LIQUID-ELEMENT THERMOCOUPLE.]
The form of thermocouple designed by the author to permit of the use of
molten elements is shown in fig. 9. A rod of refractory material, R, is
perforated longitudinally by two holes, down which are passed rods of
the thermo-elements, A and B. The lower ends of A and B are inserted in
a graphite block G, which is jointed on its upper face to R; the whole
being surrounded by the refractory cover C. On either or both of the
elements melting, the circuit is maintained through G, which serves
also to prevent the mixing of A and B when molten, whilst not affecting
the E.M.F. developed. In order to allow for the expansion of the metals
on melting, A and B are made to fit loosely in R. When inserted in a
furnace to a depth represented by EF, only the portion of the metals
adjacent to the closed end will melt, the outer parts remaining solid.
At present it has not been found possible to procure the refractory
parts in a form suited to commercial use, but when this obstacle
is overcome this type of thermocouple should prove of service for
measuring temperatures beyond the scope of ordinary base-metal
junctions.
=Indicators for Thermo-electric Pyrometers.=—As the electromotive
force developed by a single junction when heated is small, a sensitive
galvanometer is required to indicate the minute current flowing
through the circuit. Delicate millivoltmeters, of the moving-coil
type, are universally employed, as they possess the advantage of
an evenly-divided scale combined with the requisite degree of
sensitiveness. The original d’Arsonval galvanometer, consisting of a
coil suspended by a metallic strip between the poles of a horse-shoe
magnet, was used by Le Chatelier, who, by its aid, was enabled to
lay the foundations of this branch of pyrometry. Three forms of this
instrument are now in use, viz. (_a_) the suspended coil “mirror” type;
(_b_) the suspended coil “pointer” type; and (_c_) the pivoted type.
Examples of each will now be described.
[Illustration: FIG. 10.—HOLDEN-D’ARSONVAL MIRROR GALVANOMETER.]
Fig. 10 represents a mirror galvanometer working on the d’Arsonval
principle, designed by Gen. Holden, F.R.S. The horse-shoe magnet
is laminated, and an iron core, supported by a pillar, is placed
between the poles. The coil, which moves in the space between the
core and the poles of the magnet, is suspended by a thin, flat strip
of phosphor-bronze, which carries a small circular mirror. A similar
phosphor-bronze strip is fastened to the lower part of the coil,
and is continued to an adjusting-screw in the base. The ends of the
suspension strips communicate with the terminals of the galvanometer,
and a current entering at one terminal passes through the metallic
suspensions and the coil to the other. The effect of passing a current
through the coil, which is located in a powerful magnetic field is to
produce an axial movement tending to twist the suspension strips, which
movement is greatly magnified by a spot of light reflected from the
mirror on to a distant scale. When the current ceases, the untwisting
of the strip restores the coil to its former position. Galvanometers of
this type are remarkably “dead-beat” in action, that is, the movement
and restoration of the coil are accomplished without vibration. A
semi-transparent scale, placed at 1 metre distance, and 50 centimetres
long, is suitable for use with this galvanometer. When used in
workshops, it is necessary to protect a mirror galvanometer from the
vibrations produced by machinery, which would cause the spot of light
to become unsteady. The best method of effecting this is shown in fig.
11, which represents the mode of suspension devised by W. J. Lambert
for use in the Royal Gun Factory, Woolwich Arsenal. The usual supports
of the galvanometer are abolished, and the instrument suspended from
the ring of a brass tripod, so as to keep three springs partly in
compression. When suspended in this manner, a mirror galvanometer
is quite suited to commercial use; in the quiet of the laboratory
the ordinary supports may be employed. The advantage gained by using
the mirror type is that a much longer scale is possible than with
instruments furnished with a pointer, and hence greater accuracy in
determining temperature readings may be secured.
[Illustration: FIG. 11.—LAMBERT’S ANTI-VIBRATION STAND FOR
GALVANOMETERS.]
[Illustration: FIG. 12.—SIEMENS’ THERMO-ELECTRIC INDICATOR.]
In suspended coil instruments furnished with a pointer, the
construction differs only in detail from the foregoing. In place of
the mirror, a light pointer is attached to the suspension so as to
rest on the coil and a scale is furnished over which the pointer
moves. Fig. 12 is an example of this type, made by Messrs Siemens, the
suspension being contained in the tube which rises from the body of the
instrument. The maximum length of scale moved over by the extremity
of the pointer is about 6 inches, as a longer and therefore heavier
pointer would reduce the sensitiveness below the point requisite for
thermo-electric work.
In the double-pivoted type, the suspension is eliminated, and pivots
are fastened to each end of the moving coil which rest in bearings.
The turning of the coil is made to compress a hair spring, made of
phosphor-bronze; and when the current ceases the unwinding of this
spring restores the coil to its former position. The coil carries
a pointer which moves over a scale. These instruments are not so
sensitive as those in which the coil is suspended, but can be made
sufficiently sensitive to work with any kind of junction in practical
use. The pivoted form is cheaper and stronger than the suspended type,
and is used whenever sufficiently sensitive.
The “Uni-pivot” galvanometer, made by R. W. Paul, is shown in figs. 13
and 17. The coil, which carries the pointer, is circular, and moves
round a spherical core of iron placed between the poles of the magnet.
A hole is drilled in the iron core, and the coil rests on a single
bearing at the bottom of this hole. A phosphor-bronze control-spring
serves to restore the coil to the zero position. The lessened friction
due to the use of a single pivot enables this instrument to be made
very sensitive when needed, so that a relatively small rise in the
temperature of a junction may cause the pointer to traverse the whole
length of the scale.
[Illustration: FIG. 13.—PRINCIPLE OF UNI-PIVOT GALVANOMETER.]
=Special Features of Indicators.=—All moving-coil instruments,
whether suspended or pivoted, are liable to alteration of the zero
point owing to what is termed “creep.” The suspension strip, when first
fixed in position, generally possesses a certain amount of initial
torsion, which comes into operation gradually and causes a slight
movement of the coil. Similarly, in a pivoted instrument, the strength
or shape of the control-spring undergoes a gradual alteration at first,
causing the pointer to move away from the zero position. For this
reason adjusting arrangements are fitted by means of which the spot of
light or pointer may be brought back to the zero. This creeping ceases
after a time—often requiring twelve months—and if not subjected to
any strain, error from this cause does not recur to any notable extent.
With a mirror galvanometer it is better to move the scale, or turn the
galvanometer round on its axis to restore the correct zero, rather than
to twist the coil back; but with a fixed scale and pointer the only
remedy is to turn the coil bodily round. In a single-pivot indicator
constantly used in the author’s laboratory, the creep amounted to a
movement of the end of the pointer through an angle of 2 degrees in
the first few months, since when, after the lapse of several years, no
further alteration has occurred. It is advisable to test the zero point
of an indicator from time to time by breaking the circuit, and if an
error be discovered the pointer should be re-set, or an allowance made
in taking a reading.
The resistance of an indicator should be so high that the readings
should not be perceptibly altered by any fluctuations in the resistance
of the circuit which may arise in practice. If leads of considerable
length were used to connect the pyrometer with the indicator, and were
subject to fairly large alterations of temperature, the consequent
changes in the resistance of such leads would be noticeable on a
low-resistance indicator; and similarly, if a pyrometer were inserted
at different depths in a furnace at separate times, thus heating up
varying lengths of the junction wires, a discrepancy would arise for
the same reason. The resistance of an indicator, however, cannot be
raised beyond a certain point without reducing the sensitiveness below
the required limit. A mirror galvanometer of the type described may
have a resistance—partly in the coil and partly in an added series
resistance—of 1000 ohms or more, and still be sufficiently sensitive;
and in the latest types of instruments provided with pointers the
resistance may be made as high as 1000 ohms, although it is more
usually 400 to 500 ohms. Many indicators are in use, however, in which
the resistance is 100 ohms or less. As, from Ohm’s law, the current
varies inversely as the total resistance in the circuit, any alteration
in resistance should be small relatively to the total to render the
error negligible. This point is made clear in the following example:—
_Example._—A thermocouple and leads have a resistance of 5
ohms and are subject to alterations amounting to 1 ohm. To
find the errors resulting when indicators of resistances
800, 400, and 50 ohms respectively are used.
From Ohm’s law, C = E/R, the variation in C, with E
constant, will be 1 in 805, 1 in 405, and 1 in 55
respectively. As the indications are proportional to the
current, the alterations caused will be approximately
⅛ per cent., ¼ per cent., and 2 per cent. The first two
may be ignored; the last may be quite serious and lead to
the failure of an operation.
It will be seen from the foregoing that low-resistance indicators
should only be used for fixed thermocouples and short leads not subject
to temperature changes, or, in other words, in a circuit of fixed
resistance.
The resistance of an indicator, when unknown, may be found by the
following method, suggested by the author:—A resistance box is joined
at one end to one terminal of the indicator. To the other terminal a
fairly stout iron wire, 18 inches long, is connected, and a similar
length of constantan wire is coupled to the other end of the resistance
box. The free ends of the wires are twisted into a junction which is
dipped into boiling water. The deflection obtained with no resistance
in the box (D_{1}) is noted, and resistances (R) are then unplugged
until the deflection (D_{2}) is approximately one-half of D_{1}. The
resistance (G) of the indicator, ignoring that of the wires, is then
given by the formula
D_{2}R
G = ───────────────
(D_{1} - D_{2})
as may readily be proved from Ohm’s law, E being constant. This method
is extremely simple and reasonably accurate.
Reliable indicators are now procurable from many instrument-makers at
a comparatively small cost, progress in this direction having been
most marked in recent years, particularly in the case of pivoted
instruments. The most convenient form for workshop use is made with
an edgewise scale (fig. 14) and may be placed in a suitable position
fixed to a bracket. The flat-scale pattern is preferable for use on a
laboratory table, or for a portable pyrometer. The sector pattern is
also good for workshop use, the dial being visible from a distance.
[Illustration: FIG. 14.—INDICATOR WITH EDGEWISE SCALE.]
=Standardizing of Indicators to read Temperatures directly.=—The
temperature scale of an indicator, for use with a given thermal couple,
is always marked by the maker in the case of instruments furnished
with a pointer, and, generally speaking, is correct within reasonable
limits. It is customary and necessary to send with the instrument a
statement of the cold-junction temperature for which the markings
are correct; say 20° C. or 60° F. The user should then endeavour to
maintain the cold junction at this specified temperature when taking
a reading, or otherwise a considerable error may be introduced. It is
highly desirable, however, that the user should be able to perform the
standardizing himself, if only for checking purposes; and when using a
mirror galvanometer as indicator it is necessary to standardize on the
spot at which the instrument is fixed. Ability to prepare a temperature
scale is further useful, inasmuch as any good millivoltmeter, of range
0 to 20 millivolts, may be used for thermo-electric work of all kinds,
and may be calibrated for different junctions, a suitable series
resistance being added to enable E.M.F.’s higher than 20 millivolts to
be measured. Such an instrument may thus be made extremely useful, both
in the workshop and laboratory.
Standardization may be effected either by subjecting the hot
junction to several known temperatures, and noticing the deflections
corresponding thereto; or by measuring the electromotive force
developed by the junction, and calculating the corresponding
temperature from a formula which is known to hold for the range
comprehended by the instrument. The former method is simpler; and if
carefully conducted is quite accurate. The latter method possesses the
advantage that readings in millivolts may be translated directly into
temperatures when the constants of a given thermal couple are known.
It is now usual to mark indicators with a double scale, one reading
millivolts and the other temperatures.
=Standardization by Fixed Points.=—Taking any millivoltmeter
which, with a maximum of 20 millivolts at the terminals, will give a
full scale deflection, the first step is to arrange that the pointer
(or spot of light) shall just remain on the scale at the highest
temperature to be attained by the junction. This may be done by placing
the hot junction in boiling water and noting the deflection obtained,
either in millivolts or equal arbitrary divisions, and also the
temperature of the cold junction. The deflection observed is due to
a difference of temperature (100-_t_) deg. C, where _t_ is the
temperature of the cold junction. If the highest temperature to be
measured is 10 times (100-_t_), the deflection should be rather
less than 1/10 of the scale, and similarly for any other required
temperature limit. If the observed deflection exceed this proportion, a
series resistance should be added until the correct value is obtained.
This resistance is then permanently installed in the circuit for use
with the junction under trial.
Before proceeding further it is necessary to consider whether the
pyrometer is to possess a single cold junction of ascertainable
temperature (as in fig. 6), or whether it will be arranged with two
cold junctions in the head, as in fig. 4. In the former case it is
simpler to prepare a “difference” scale; that is, one which reads
differences of temperature between the hot and cold junctions, from
which the temperature of the hot end may be obtained by adding to the
difference that of the cold junction. In the latter case the cold end
should be kept by artificial means at the temperature likely to be
attained in practice—say 25° C.—a water-bath being suitable for this
purpose. It is advisable to remove the shield of the pyrometer when
standardizing, so as to expose the hot junction, as closer readings can
then be taken.
A number of materials—preferably cheap—of known boiling points or
melting points are then selected from a table of fixed points (page
16) so as to give about six points, distributed fairly evenly over the
scale. As an example, if it were desired to prepare a temperature scale
from 0° to 1000° C., the following might be chosen:—
════════════════════════════════════════════════╤═══════════════════
Substance and Condition. │ Temperature.
────────────────────────────────────────────────┼─────────┬─────────
Water at boiling point │ 100° C. │ 212° F.
Tin at melting point │ 232 │ 449
Zinc at melting point │ 419 │ 786
Antimony at melting point │ 631 │ 1167
Common salt at melting point │ 800 │ 1472
Copper at melting point (covered with graphite) │1084 │ 1983
════════════════════════════════════════════════╧═════════╧═════════
The hot junction is allowed to attain these temperatures successively,
and the corresponding deflection in each case is noted. It is then
possible to divide up the whole of the scale to read temperatures
directly.
The first reading is taken by placing the junction in a vessel of
boiling water, and for a locality near sea level it is not necessary
in ordinary work to take account of fluctuations in the boiling point
due to alterations of atmospheric pressure. To ensure that the other
readings are taken when the substances are exactly at the melting
point, the procedure is as follows: about 2-3 lb. of the substance are
melted in a salamander crucible, and a small fireclay tube, closed at
one end, is inserted in the molten mass. The hot junction is placed
in the fireclay tube, and the intervening space filled with asbestos
fibre. Great care must be taken not to let the junction touch the fused
substance. The crucible is now allowed to cool, and a reading of the
deflection taken every half-minute. When the substance is exactly at
its solidifying point—identical in general with the melting point—the
deflection remains stationary for several consecutive readings, owing
to the liberation of latent heat of fusion in sufficient quantity
to balance the loss by radiation. This stationary reading is noted
for each substance, and represents the deflection given when the hot
junction is at the temperature corresponding to the melting point, and
the cold junction or junctions at the temperature existing when the
observation is made. For melting the materials, a Davies furnace with
a large Teclu or Meker burner is convenient up to 850° C.; but to melt
the copper a blast lamp is requisite. The molten mass may be allowed to
cool in the furnace.
From these observations a calibration curve may be drawn either for
differences between hot and cold junctions, or for a steady temperature
of the cold junctions. Two sets of data are appended to illustrate the
procedure.
════════════╤═════════════════════════════════╤═════════════════════
Temperature │ Pyrometer 1. Iron-constantan. │ Pyrometer 2.
of Hot │ (Series resistance in │ Platinum-
Junction. │ galvanometer circuit.) │ iridioplatinum.
├───────────┬─────────┬───────────┼───────────┬─────────
│Deflection.│ Cold │Difference.│Deflection.│ Cold
│ │Junction.│ │ │Junction.
────────────┼───────────┼─────────┼───────────┼───────────┼─────────
100° C. │ 8·9 │ 15° C.│ 85° C. │ 5·5 │}
232 │ 21·8 │ 17 │ 215 │ 15·6 │}Constant
419 │ 40·6 │ 19 │ 400 │ 29·4 │} at
631 │ 63·8 │ 19 │ 612 │ 45·5 │} 25° C.
800 │ 83·0 │ 20 │ 780 │ 59·0 │}
1084 │ ... │ ... │ ... │ 82·0 │}
════════════╧═══════════╧═════════╧═══════════╧═══════════╧═════════
[Illustration: FIG. 15.—CALIBRATION CURVES FOR TWO THERMO-ELECTRIC
PYROMETERS.]
Fig. 15, A, is a calibration curve for thermocouple 1, connecting
deflections with corresponding differences between the temperatures
of the hot and cold junctions. In order to read from this curve the
temperature of the hot end, the reading corresponding to the observed
deflection is added to the existing temperature of the cold junction.
Thus if a deflection of 56 divisions were obtained with the cold
junction at 25°, the temperature of the hot junction would be (540 +
25) = 565° C. The advantage of this method of calibration is that it is
unnecessary to take precautions to keep the cold junction at a steady
temperature; and when a single cold junction is used, as in fig. 6,
this plan should always be followed. It will be noted that this curve
passes through zero, as no deflection represents no difference of
temperature.
Fig. 15, B, represents the calibration curve for pyrometer 2, and
is such that direct readings may be obtained corresponding to any
given deflection, for a cold junction temperature of 25°. This curve,
therefore, cuts the axis of zero deflection at 25°, as no deflection
corresponds to the condition when both hot and cold junctions are at
25°. This method of calibration may be used with advantage for couples
of the type shown in fig. 4, where two cold junctions exist in the
head, and the simple rule of adding the cold junction temperature
does not apply. Many suggestions have been made for correcting for
alterations in the temperature of the cold end of such a couple,
but none are accurate, and it is necessary to keep this part at the
temperature of standardization to secure correct readings. In both of
the above calibrations the galvanometer used possessed a scale divided
into 100 equal arbitrary divisions.
In making permanent temperature scales from these curves to attach
to the existing galvanometer scale, intervals of 100° may be taken
and marked opposite to the corresponding divisions on the existing
scale. Each 100° may then be equally subdivided into as many parts as
the length of scale permits, and numbered at suitable intervals. If
the junction used yield a calibration curve departing greatly from a
straight line, every 50° interval should be taken, or, if necessary,
every 25°. In the examples given both curves are nearly straight lines
in the working region, viz. 400° to 800° for the iron-constantan
junction, and 500° to 1100° for the platinum-iridioplatinum.
One precaution necessary in standardizing an indicator by this method
is to ensure that the metals used are pure, as impurities lower the
melting points. If ordered as “pure” from any dealer of repute, the
metals will generally be found satisfactory. The common salt used
should be the ordinary salt sold in blocks, and not a prepared table
salt. A second precaution, when observing melting points, is to guard
against a possible error due to the substance becoming “surfused” or
“overcooled”; in which case the temperature falls below the ordinary
freezing point before solidification commences. When freezing occurs,
however, the temperature rises to and remains at the true melting
point, and an increase of deflection following a gradual fall always
indicates overcooling. The higher deflection then attained is the true
freezing point. Antimony frequently overcools to 600° before freezing,
but on setting rises to the correct figure—631°. All metals and salts
are liable to overcooling occasionally.
=Standardization by Measurement of E.M.F.=—It has been found,
as the result of experiments, that the relation between the E.M.F.
developed by a junction and its temperature—under constant conditions
of the cold junction—may be expressed approximately by a formula as
under:—
log E = A log _t_ + B (Holman’s formula),
where E = electromotive force in microvolts, _t_ = temperature in
Centigrade degrees, and A and B are constants depending upon the
junction. With certain junctions this formula may be applied over
the working part of the scale with an error not exceeding 2° C., but
with others the discrepancy is greater. In order to determine the
constants A and B, it is necessary to measure the E.M.F. at two known
temperatures, which should be chosen as far apart as possible in the
working region. When these constants are known, a measurement of E
enables the temperature _t_ to be found by calculation.
_Example._—Le Chatelier found that a junction at the
temperature of melted aluminium (657° C.) gave 6200
microvolts; at the melting point of copper in air (1062° C.)
the figure was 10580. Applying in the above formula
log 6200 = A log 657 + B
and log 10580 = A log 1062 + B,
the value of A is 1·2196 and of B 0·302, as may be found
by taking logarithms and solving for A and B.
The values of the constants A and B vary for different junctions, and
also for different melts of what are reputed to be the same materials.
When once determined for a quantity of homogeneous wires, to which
the formula applies with sufficient accuracy, it is evident that an
indicator with a millivolt scale may be made to read temperatures
directly without any necessity for further experiment, although it is
always advisable to take one check reading at a fixed point in the
working range.
[Illustration: FIG. 16.—POTENTIOMETER METHOD OF MEASURING E.M.F.]
In order to determine the E.M.F. of a junction at different
temperatures, the potentiometer method is used, in which the E.M.F.
of the test-couple is balanced against the known E.M.F. furnished by
a constant cell. The circuit is shown in fig. 16, in which B is an
accumulator which sends a current through the resistances R_{1}, R_{2},
and the calibrated wire DE. The cold ends of the couple are attached
at P so as to be in opposition to B, and in this branch of the circuit
are included a sensitive galvanometer G and a portion of the wire DE. A
standard cadmium cell, S, is connected between R_{1} and R_{2} at one
end, and may be put in circuit with the galvanometer through the switch
A. In commencing, S is connected to the galvanometer and R_{1} adjusted
until no deflection is obtained on G. The switch A is now moved over
to the circuit of the couple, and the terminal F moved along the wire
until zero deflection is again obtained. The E.M.F. of the couple is
determined from the relation
E of junction Resistance of DF
────────────────── = ───────────────── .
E of standard cell R_{2}
By exposing the hot end of the junction to successive standard
temperatures, and maintaining the cold ends at a known constant
temperature, the necessary data for inclusion in a formula may be
obtained.
In fixing a permanent temperature scale, calculated from the formula,
to a millivoltmeter, it must be remembered that the values given by
the experiment are absolute, and independent of the resistance of the
circuit composed of the thermo-element and galvanometer. On the other
hand, a millivoltmeter is marked to read difference of potential at
its terminals; and if in series with a junction and leads of notable
resistance, its indications will not be the E.M.F. of the junction. An
example will make this point clear.
_Example._—A millivoltmeter has a resistance of 100
ohms, and is marked to read P.D. at its terminals. A
thermocouple and leads connected to the millivoltmeter
have a resistance of 5 ohms. To find the relation between
the true E.M.F. of the junction and the readings of the
indicator.
If E = the E.M.F. developed by the junction, and V, the
reading of the millivoltmeter, = P.D. at its terminals,
then the current in the circuit = E/105 = V/100; and V =
(100/105)E. That is, the readings are lower by 5 per cent.
than the true E.M.F. of the junction. In the same way a
low resistance voltmeter, if applied to a cell of high
resistance, shows a lower reading than the E.M.F. of the
cell.
This example indicates how a table connecting true E.M.F.’s with
reading in millivolts may be calculated when the resistances concerned
are known. It is presumed, in preparing a scale in this manner, that
the resistance of the couple will not be subject to such alterations as
to affect the reading.
The advantages of this method of calibration are manifest when a number
of junctions are being made from a given batch of wires, as it is only
necessary to divide the scale of the indicator so as to represent
millivolts—a simple operation—and then to attach a temperature scale.
This procedure is much more expeditious than standardizing each
indicator at several fixed points when a number are concerned, but for
a single junction the fixed point method is easier. The potentiometer
method of measuring E.M.F. may also be used to determine temperatures
in place of an indicator, and is of great service in cases where very
accurate readings are specially required, being far more delicate in
detecting small differences of temperature than an indicator. Special
potentiometers for thermo-electric work are made by the Cambridge
and Paul Instrument Company, Siemens, and others, and are useful in
conducting accurate research, but are too elaborate for workshop or
ordinary laboratory practice.
=Cold Junction Compensators.=—The necessity for paying attention
to the cold junction has led to various attempts to compensate
automatically for changes of temperature at this part of the pyrometer.
A thermometer located near the cold junction, as in fig. 6, is all that
is needed to correct a two-junction circuit; but when a three-junction
circuit is used a correct reading is not secured by adding the excess
temperature of the thermometer over the calibration temperature to
the reading on the indicator. In Bristol’s arrangement a mercury
thermometer, with a large bulb and wide stem, is stationed at the cold
junction, and participates in any temperature change. In the stem is
placed a loop of thin platinum wire, which forms part of the pyrometer
circuit. When the mercury is heated it expands up the stem and
short-circuits a portion of the loop, thereby diminishing the
resistance of the pyrometer circuit, and tending to increase the
deflection on the indicator. Simultaneously the cold junction will
be heated, tending to diminish the current, and so to cause a less
deflection. By adjustment these two tendencies may be counterbalanced,
so that the reading is unaffected, but such adjustment will only
apply to a given E.M.F., and therefore to one temperature of the hot
junction. Hence this method fails in general application.
Peake’s compensated leads are intended to remedy cold-junction errors
by transferring this junction, in effect, to the galvanometer. They
are used for pyrometers in which the platinum metals are employed, and
consist of wires of two different alloys of copper and nickel, which
connect the cold end to the indicator. These alloys are such that
the electromotive forces set up at the junctions in the head—Pt and
Cu-Ni 1, and Pt-Ir with Cu-Ni 2—are equal and opposite at all working
temperatures, and hence changes at the cold junctions do not affect the
reading. At the indicator, however, temperature changes would cause an
alteration in deflection; but as the indicator is generally placed well
away from the furnace, and is not liable to notable heating or cooling,
the possible errors are greatly reduced by the use of these leads. They
are obviously of no value for use with base-metal pyrometers, as the
wires used in such may be prolonged to the indicator, with an identical
result.
[Illustration: FIG. 17.—DARLING’S COMPENSATOR, FITTED TO GALVANOMETER.]
An automatic compensator for use with base-metal pyrometers has been
devised by the author, and is illustrated in figs. 17 and 18. A spiral
made of a compound strip of two metals is attached to the needle of the
indicator, and coils or uncoils when cooled or heated, thereby moving
the pointer over the scale. The length of the spiral is such that an
alteration of a given number of degrees in its temperature moves the
pointer by the same number of degrees on the scale—or, in other words,
the temperature scale of the pyrometer is identical with that of the
spiral. The metals forming the junction are continued, in the form of
wires, to the interior of the galvanometer, where a cold junction is
formed, which will always possess the same temperature as the spiral.
The scale is constructed to represent differences of temperature between
the hot and cold junctions, and before coupling up the pyrometer the
pointer indicates the temperature of the spiral; that is, of the cold
junction. On connecting the thermocouple the pointer is moved by the
coil of the indicator through an amount represented by the difference
in temperature between the two junctions, and therefore finally
indicates the temperature of the hot junction.
[Illustration: FIG. 18.—INDICATOR FITTED WITH DARLING’S COMPENSATOR.]
_Example._—If the cold junction were at 20°, the pointer,
before connecting the couple, would indicate 20° on the
scale. If the hot junction were 580° hotter than the cold,
then on completing the circuit the pointer would move 580
additional degrees along the scale, so that the figure
indicated would be (20 + 580) = 600°, the temperature of
the hot junction. If now the indicator were heated by 10°,
the spiral would tend to augment the deflection by 10°,
but simultaneously the deflection due to the junctions
would fall off by 10°, and the reading would still be 600°.
This method of compensation renders the readings independent of the
cold junction, and, in addition to its use for high temperatures,
enables ordinary and low temperatures to be read simply and correctly,
as will be shown later. The spiral is located in the tower rising from
the top of the indicator in fig. 18.
In Paul’s method of compensation the thermocouple and indicator
are placed across a Wheatstone bridge, two arms of which contain
resistances of copper, whilst the resistances in the other two arms are
of manganin. Any change in temperature at the cold junction is shared
by these four resistances, and, whilst affecting the resistance of the
copper parts, no change is caused in the manganin parts, as this alloy
has a negligible temperature coefficient. If, therefore, the bridge
were initially balanced at 20° C., and the temperature rose to 30°,
the increased resistance of the copper would destroy the balance, and
permit of a small current passing through the indicator. A fall to
10°, by diminishing the resistance of the copper, would cause an equal
current to pass through the indicator in the opposite direction. The
amount of this current is arranged so as to add the rise in temperature
of the cold junction to the reading of the indicator in the one case,
and to subtract the fall in the other, thus retaining true readings for
the cold-junction temperature at which the couple was standardized.
=Constant Temperature Cold Junctions.=—If the cold junction
can be kept at a steady temperature, compensators are unnecessary,
but no good practical means of achieving this end has yet been
devised. Water-cooled heads have already been referred to; but in many
situations the connecting-pipes entailed would be objectionable, and
hence this arrangement is not greatly used. An alternative method,
suggested by Prof. A. Zeleny, is to bury the cold junction in the
ground. Recent experiments, conducted at Cambridge by R. S. Whipple,
showed that a junction buried 10 feet deep did not vary in temperature
by more than 2° C. over a period of three years. This has led to the
adoption of buried junctions in special cases; but it is probable that
much greater variations would be experienced in the ground beneath
large furnaces, in which case the advantages of this procedure would
be lost. A common workshop method is to locate the cold junction in a
thermos flask filled with oil, when a temperature constant to 2° C. may
be secured, although the changes in the temperature of the surrounding
atmosphere may be as great as 150 C. For special work, ice may be
used in the thermos flask, thus securing absolute constancy; but this
procedure is not feasible in ordinary works practice.
=Special-Range Indicators.=—When the working range of a pyrometer
is from 600° C. upwards, it is evident that the part of the scale
occupied by the first 600° is useless, and that it would be an
advantage if the whole scale could be utilised for the special working
range, so as to secure more exact readings. This may be accomplished
by a “set-up” against the movement of the pointer caused by the
thermocouple, so as to prevent any motion over the scale until an
assigned temperature is reached. For example, a junction developing
12 millivolts at 1000° C. may be coupled to an indicator in which the
full-scale deflection of the pointer is produced by 6 millivolts. If an
E.M.F. of 6 millivolts be opposed to the junction, no deflection will
occur until the temperature at which the couple develops 6 millivolts
is reached—when the opposing E.M.F. will be overcome. This temperature
may be 500° C., so that the whole scale may be divided up between
500° and 1000°. The length of the indicator scale is thus effectively
doubled; and by using different values for the set-up, it is evident
that any desired range may be obtained within the limits of sensitivity
of the indicator. The method of procuring the opposing E.M.F. varies
with different makers. The Cambridge and Paul Instrument Company employ
a dry cell and a series resistance, connected so as to oppose the
thermocouple; and by adjusting the resistance any desired set-up may be
obtained, the value of which, in degrees, may be read off by connecting
the cell and resistance to the indicator, the couple having been
switched out of the circuit. Thus, to adjust for a range of 500°-1000°
on an indicator giving full-scale deflection for 500°, the resistance
is regulated so that the cell alone causes the pointer to move to the
end of the scale. The method adopted by Paul consists of suitable
resistances inserted in a Wheatstone bridge, which may be thrown off
the balance, and thus cause an opposing E.M.F. of the correct amount at
the terminals of the indicator.
A mechanical set-up has been introduced by the Cambridge and Paul
Instrument Company, the indicator in this case having a suspended
coil. By turning a milled-head a twist may be given to the suspending
strip, and by the turning of a second head the pointer may be brought
back to zero, retaining the initial twist, which is opposed to that
produced by the current due to the couple. Thus, if the imposed twist
were such as to move the pointer to the 400° mark on the scale, the
temperature indicated by the junction would be the observed reading
plus 400. By this method it is possible to obtain any desired range
within the limits of the indicator. The danger of producing errors due
to “creeping” is said to be negligible.
[Illustration: FIG. 19.—CIRCUIT OF NORTHRUP’S “PYROVOLTER.”]
=Potentiometer Indicators.=—The advantage of measuring E.M.F.
by the potentiometer method is that the result is independent of the
resistance of the circuit under test, whereas an indicator is affected
by changes in the resistance of the circuit in which it is inserted.
When long leads are used to connect a couple to its indicator, notable
errors may be caused by the varying resistance of the leads, due
to changing temperature; and, in addition, the resistance of the
couple-wires varies according to temperature and depth of insertion in
the furnace. Attempts have therefore been made to produce indicators
based on the potentiometer principle, suitable for workshop use, and
one form, known as Northrup’s “Pyrovolter,” is arranged as shown in
fig. 19, A. A cell D sends a current through a rheostat R, a copper
coil C, and a manganin coil S. The copper coil has the same resistance
as the copper winding of the indicator G. The couple is connected,
with G in circuit, across the manganin coil S, the resistance of this
material being unaffected by temperature. By adjusting R until no
deflection is shown on G, the drop of volts across S is made equal
to the E.M.F. of the couple. To measure this drop, a key is pressed,
altering the circuit as shown in B, the indicator being now in series
with S and the couple detached. The value of the current passing
through S is unchanged, as the indicator coil has the same resistance
as the copper coil C, which it now replaces. The deflection on G
indicates the value of this current, and, as the drop of volts across
S is proportional to the current, G may be marked off to read E.M.F.
and the corresponding temperature of the junction. The advantages
claimed are that the indicator may be used with any type of junction,
and is unaffected by temperature changes in the circuit. A similar
instrument is made by the Brown Company of Philadelphia. Up to the
present potentiometer indicators have not been adopted to any extent
in Britain, and the adjustments necessary to obtain a reading must be
accounted a distinct drawback from a workshop standpoint.
=Recorders for Thermo-electric Pyrometers.=—It is frequently of
importance to know not only the existing temperature of a furnace, but
also the fluctuations to which it is subject. Continuous observation of
a pyrometer would involve too much labour, and it is therefore evident
that an automatic recorder would possess many advantages in such cases.
A continuous record shows whether the attendant has maintained the
temperature between the prescribed limits, and furnishes a permanent
history of a given operation, which often serves as a guide to future
procedure.
The first successful recorder, suggested by Sir W. Roberts-Austen and
designed by Gen. Holden, F.R.S., was used in conjunction with a mirror
galvanometer. In its original form, the spot of light from the mirror
was made to fall on a sensitized plate, to which a gradual vertical
motion was conveyed by connecting the dark slide to a water-float by
means of a chain and pulley. The float was placed in a tank of water,
which was gradually emptied through a tap, causing the float to sink
and the plate to rise. If the deflection of the spot of light remained
steady, a vertical straight line was traced on the plate, fluctuations
producing a sinuous line. Trials at known temperatures enabled a
standard plate to be obtained, divided into degrees, which could be
superposed on a trial plate, and the temperatures thus determined. Much
valuable work was accomplished with this recorder by Roberts-Austen
for the Alloys Research Committee of the Institution of Mechanical
Engineers.
[Illustration: FIG. 20.—ROBERTS-AUSTEN RECORDER.]
In its modern form (fig. 20) the photographic plate is replaced by a
sheet of sensitized paper wound round a drum which rotates at a known
rate—say, once in 12 hours—by means of internal clockwork, shown to
the left of the figure. The galvanometer is placed at the opposite
end, and the mirror is illuminated by means of an electric lamp placed
externally, the rays from which are reflected from a prism in the
interior on to the mirror. The ray of light leaving the mirror is
broken into two portions, one of which passes through a narrow slit on
to the sensitized paper, whilst the other portion is reflected on to
a ground-glass scale on the lid, divided so as to read temperatures.
In this manner the arrangement serves not only as a recorder, but
also indicates the existing temperature without necessitating the
examination of the sensitized paper. The whole arrangement is made
impervious to light, so that it may be used in daylight. A dark room is
necessary for fixing the records. When desired, records of two or more
pyrometers may be taken on the same sheet, a clockwork device being
used to switch each instrument in turn on to the galvanometer for a
given period, an external dial indicating which pyrometer is for the
time being in circuit.
Whilst it is a drawback to the use of this recorder that the record is
not visible, the use of a mirror galvanometer confers a high degree of
sensitiveness to the instrument, not possessed by the recorders to be
described subsequently.
[Illustration: FIG. 21.—PRINCIPLE OF THREAD RECORDER.]
=The Thread Recorder.=—In this instrument an intermittent record
is secured in ink, possessing the advantages of visibility during
the period over which readings are taken, and of permanence without
subsequent treatment of the chart. The principle is shown in fig.
21, where A is a boom terminating in a V-shaped piece of ivory, and
attached to the galvanometer suspension B. By means of a cam E, rotated
by clockwork, a bar D is made to descend at stated intervals, pressing
the end of A on to an inked thread G, and causing the thread to touch
a paper wound round the drum C. This drum rotates on its axis once in
25 hours by the action of internal clockwork. The continued rotation
of the cam E alternately raises and depresses the boom A, leaving it
free for a sufficient time to enable it to attain the position it would
occupy if the mechanism were absent. The thread G is passed over
pulleys, and is wound round through an ink-well, so that the portion
opposite A is always moist. With the bar D descending every two
minutes, the successive dots form a nearly continuous line. The paper
on C is divided horizontally into temperatures, and vertically into
time units, so that the temperature existing at any given time may
readily be ascertained. The front of the bar D, or a separate strip
parallel to it, is divided so as to enable temperatures to be read
without reference to the chart. The actual instrument is shown in fig.
22. When several simultaneous records are required, the drum C is
extended, and other galvanometers introduced, to which the separate
pyrometers are connected. Several records can be taken on one chart by
introducing a clockwork mechanism to couple each pyrometer in turn to
the one galvanometer.
[Illustration: FIG. 22.—THREAD RECORDER.]
[Illustration: FIG. 23.—SIEMENS’ RECORDER.]
=The Siemens Recorder.=—In this instrument (fig. 23) the boom
from the galvanometer terminates in a knife-edge, and moves over a
thin horizontal rail, the top of which is rounded. Between the rail
and the boom are placed an inking ribbon and a paper chart, which is
moved forward by clockwork. A chopper-bar, also actuated by clockwork,
descends at about half-minute intervals, and depresses the end of the
galvanometer boom, thus producing a small dot on the chart. The paper
is 12 cms. wide and 40 yards long; it is divided into time and
temperature units, and moves forward at the rate of 2 cms. per hour.
Levelling screws are fixed to the base of the recorder.
[Illustration: FIG. 24.—FOSTER’S RECORDER.]
=Foster’s Recorder.=—Foster’s recorder (fig. 24) is designed
for use with base-metal couples of the nickel-chromium type, known as
Hoskin’s alloys, which yield an E.M.F. about five times as large as
a platinum-rhodioplatinum couple. The force available in this case
enables the coil of the galvanometer to be pivoted in a horizontal
position, the pointer being vertical, and yet to be sufficiently
sensitive. The chart is mounted on a vertical plate which rotates on
its axis, the time ordinates taking the form of concentric circles,
which are cut at an angle by the temperature ordinates. At the terminus
of the pointer is placed a small capillary tube, fitted with an inked
wick, which, when pressed upon the chart, makes a mark. The presser-bar
is curved to the same radius as the pointer, and carries a pad wetted
with ink, so that at each depression the supply of ink to the wick is
replenished by an amount equal to that imparted to the chart. This
recorder is sometimes fitted with special contacts, so that when the
correct temperature exists an electric lamp with a white bulb remains
lighted; whereas when too low or too high a green or red lamp is lit
up, and an alarm thus given. Such an addition involves the use of a
relay circuit, but is advisable in cases where expensive articles
might suffer if overheated. It can be modified to permit of several
simultaneous records being taken, and possesses the advantage that the
whole chart is visible at any time. On the other hand, the circular
coordinates may be accounted a drawback by some, as not being quite
s
Cite this
Charles R. Darling. "Pyrometry: A Practical Treatise on the Measurement of High Temperatures". Project Gutenberg. Public domain in the USA. https://www.gutenberg.org/ebooks/56353 JSON record · API collection · BibTeX · RIS · CSL-JSON · citation policy